17
8027092/AE00/V1-0/2022-10| SICK
O P E R A T I N G I N S T R U C T I O N S | VISIC100SF
Subject to change without notice
MENUS
6
6.7
Menu - Installation and Calibration
6.7.1
MP Parameters
Function: Display measuring point parameters
Display
Displays the measuring range of the active measuring point
Procedure:
1
Menu: Installation & Calibration
→
MP Parameter
→
DP1
→
Open Nom/Gas/Unit.
2
Press Down arrow key and display the measuring range.
6.7.2
System Parameters
Function: Display system parameters
System Parameters
●
System information: Serial number
●
Power-on time
Power-on time
Gas sensors require a warm-up time until the measurement process reaches a stable
state. During this warm-up time, the sensor signal can lead to the unwanted triggering of a
pseudo alarm. Therefore, the power-on time is started after switching on the voltage or
voltage recovery. During this time, the sensor outputs “Warm-up time” instead of the
measured value.
The factory-set power-on time is 5 min. and can only be read out.
Note: Changed parameters must be read back to check for correctness!