COMMISSIONING
6
6.3
Alignment
Fig. 21: Menu Adjustment / Alignment.
1
Open “Adjustment/Alignment” in the SOPAS menu tree.
Fine adjustment of horizontal alignment
1
Slightly loosen the retaining screws of the measuring device on the assembly console so
that the device can be moved.
2
Using a small plastic or rubber hammer, very carefully tap the position shown below (
Fig. “Vertical sensor alignment”, page 34
3
Tap until the
VIS X value
on the display reaches <0.15.
4
Tighten the retaining screws again.
The mechanical alignment of the measuring device and the reflector must have already
been performed by using the laser adjustment units.