OPERATION
Activate the QA analog output
b
Set the parameter in the QAMENU-SimCur or SimVol menu to the desired signal
value.
•
SimCur for current output
•
SimVol for voltage output
Note:
The simulation is automatically deactivated if the supply voltage is interrupted.
Simulating the level
Even if there is no liquid in the container yet, it is possible to select a fill level in the
menu in order to test the sensor configuration. When simulating a level value, all out
-
puts on the LFP are set according to the defined configuration. The function should not
be selected until a configuration is complete.
Configuration
b
Set parameter to the desired fill level as a % in the SimLev menu.
Note:
• The level simulation refers to the probe length and/or container level (probe
offset) if an offset is configured.
• The simulation is only active when there are no error messages. The simulation is
automatically deactivated if the supply voltage is interrupted.
Parameter selection
•
SimOff: Off
•
Fill level 0%
•
Fill level 25%
•
Fill level 50%
•
Fill level 75%
•
Fill level 100%
8.4.7
Configuring the probe length
1. Log in to expert mode, see
2. Access the EXPRT-CONFIG-Length menu using the arrow and Set pushbuttons.
3.
Enter the probe length in the Length menu. Please note the defined probe length in
Note:
•
HiSpd: max. length = 2,005 mm, response time < 400 ms
•
HiAcc: max. length = 6,005 mm, response time < 2,800 ms