O P E R AT I N G I N S T R U C T I O N S | L F P I N OX
8019921/15H0 / 2019-09-30| S I C K A G
Irrtümer und Änderungen vorbehalten
2 1
7
COMMISSIONING
7
Commissioning
7.1
Quick commissioning (with factory settings)
Quick commissioning is used in applications under reference conditions,
.
Where the following applies:
•
Use in metallic containers or immersion / bypass pipes
•
Situations where the liquid to be measured has a DK value of > 5; see
.
Commissioning
1. Mount the sensor as appropriate to the installation conditions, see
2. The container must be empty and/or the level must be at least 200 mm below the
end of the probe.
3. Log in to expert mode, see
4. After mounting, launch the AutCal menu item.
•
Press and hold the Set pushbutton for at least 3 secs.
•
Use the Set pushbutton to confirm the AutCal menu item and then use it again
to confirm the Ok? prompt.
•
The AutCal function is confirmed with !CalOK.
5.
Configure outputs, see
“8.2 Configuring the switching outputs”.
Note:
If the AutCal function was confirmed with !NoSig, relaunch AutCal.
For problems during commissioning, see
7.2
Advanced commissioning
Advanced commissioning is required when quick commissioning is not sufficient or if
one of the following situations applies:
•
The liquid to be measured has a DK value of < 5, see
.
•
There are extreme ripples on the surface of the liquid.
•
There are variations in the installation conditions, see
.
•
If the tank cannot be completely emptied prior to calibration for technical reasons.