DUSTHUNTER T · Operating Instructions · 8012428 V2.2 · © SICK AG
Start-up and Parameter Settings
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
4.4.5
Setting the analog inputs parameters
Select the "Configuration / I/O Configuration / Input Parameters DUSTHUNTER" directory
to set the analog inputs.
Figure 66
"Configuration / I/O Configuration / Input Parameters" directory
Field
Parameter
Remark
Temperature
Source
Constant Value
A fixed value is used to calculate the scaled value.
This parameter opens the "Temperature Constant" field to enter the scaling value in °C
(° F for imperial units) or K.
Analog Input 1
The value from an external sensor connected to analog input 1 (standard scope of delivery) is
used to calculate the scaled value.
This parameter opens the "Analog input 1 - Temperature" field to set the lower and upper
range limit values and the Live Zero value.
Pressure
Source
Constant Value
A fixed value is used to calculate the scaled value.
This parameter opens the "Constant Temperature" field to enter the scaling value in mbar
(= hPa).
Analog Input 2
The value from an external sensor connected to analog input 2 (standard scope of delivery) is
used to calculate the scaled value.
This parameter opens the "Analog input 2 - Pressure" field to set the lower and upper range
limit values and the Live Zero value.
Moisture
Source
Constant Value
A fixed value is used to calculate the scaled value.
This parameter opens the "Constant Moisture" field to enter the scaling value in %.
Analog Input 3
The value from an external sensor connected to analog input 3 (optional module required) is
used to calculate the scaled value.
This parameter opens the "Analog input 3 - Moisture" field to set the lower and upper range
limit values and the Live Zero value.
Oxygen
Source
Constant Value
A fixed value is used to calculate the scaled value.
This parameter opens the "O2 Constant Value" field to enter the scaling value in %.
Analog Input 4
The value from an external sensor connected to analog input 4 (optional module required) is
used to calculate the scaled value.
This parameter opens the "Analog input 4 - Oxygen" field to set the lower and upper range
limit values and the Live Zero value.