Maintenance
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
Remove pivoted shutter cover (1), press tension spring (2) together and take pivoted
shutter (3) off axis (4).
Carefully clean glass pane (5) (both sides), zero point reflector (6) and sender optics (7)
with an optics cloth.
Figure 86
Cleaning the optical interfaces on the sender/receiver unit
Lay the toothed belt on the drive axis, press the tension spring together and push piv-
oted shutter back onto the axis.
Start the function check by moving device file "MCU" to the "Project tree" window"(
), select the "Adjustment / Function Check Manual" subdirectory and click
the "Start Manual Function Check" button.
Figure 87
"Adjustment / Function Check - Manual" subdirectory
3
6
Zero point reflector,
rear side
1
7
5
4
2
The function check can also be triggered via the LC-Display on the MCU (