Maintenance
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
Remove pivoted shutter cover (1), loosen tension spring (2) and take pivoted shutter (3)
off axis (4).
Carefully clean glass pane (5) (both sides) and reflector optics (6) with an optics cloth.
Figure 91
Cleaning the optical interfaces on the reflector
Lay the toothed belt on the drive axis, push pivoted shutter back onto the axis and rein-
stall the tension spring.
Trigger the function check (select the "Adjustment / Function Check - Manual" subdirec-
tory in the "MCU" device file and click "Start Manual Function Check";