78
DUSTHUNTER T · Operating Instructions · 8012428 V2.2 · © SICK AG
Start-up and Parameter Settings
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
The "Analog Output 2(3) Parameter" and "Analog Output 2(3) Scaling" fields
only appear after activation of the "Use first analog output module" checkbox
and when an AO module is plugged in (standard on DUSTHUNTER T100 and
DUSTHUNTER T200).
Field
Parameter
Remark
Analog
Outputs
- General
configuration
Output Error
current
Yes
Error current is output.
No
Error current is not output.
Error Current
Value < Live Zero (LZ)
or > 20 mA
mA value to be output in "Malfunction" state (error case) (size
depends on connected evaluation system).
Current in
maintenance
User value
During "Maintenance", the value entered in the "Maintenance
current" field is output.
Value measured last
The value measured last is output during "Maintenance"
Measured value output The current measured value is output during "Maintenance".
Maintenance
current
Whenever possible,
value
≠
LZ
mA value to be output in "Maintenance" state
Optional
Analog Output
Modules
Use first analog
output module
Inactive
Not permitted for DUSTHUNTER T100/T200 (results in error, because
AO 2 and AO 3 are available by default).
Active
Opens the fields to set parameters for AO 2 and AO 3 (standard on
DUSTHUNTER T100 and T200)
Analog
Output 1
Parameter
Value on analog
output 1
Concentration a.c.
(Ext)
Dust concentration under operating
(actual) conditions (based on
extinction)
The selected measured
variables are output on the
analog output.
Concentration s.c. (Ext) Dust concentration under standard
conditions (based on extinction)
Opacity
Extinction
Transmission
Rel. opacity
Relative opacity
Live zero
Zero point
(0, 2 or 4 mA)
Select 2 or 4 mA to ensure being able to differentiate between
measured value and switched off device or interrupted current loop.
Output
checkcycle
results on the AO
Inactive
Control values (
p. 15, §2.1.3) are not output on the analog output .
Active
Control values are output on the analog output (the "Output control
values at AO" checkbox in the "Adjustment / Function Check -
Automatic" directory must be activated).
Write absolute
value
Inactive
Positive and negative measured values are differentiated.
Active
The amount of the measured value is output.
Analog
Output 1
Scaling
Range low
Lower measuring
range limit
Physical value at live zero
Range high
Upper measuring
range limit
Physical value at 20 mA