MAINTENANCE AND CARE
9
Maintenance and care
Claims under the warranty rendered void!
The housing screws of the Bulkscan are sealed. Claims under the warranty against
SICK AG will be rendered void if the seals are damaged or the sensor opened. The
housing must only opened by SICK authorized service personnel.
9.1
Cleaning the front screen
The Bulkscan is largely maintenance-free. The front screen on the Bulkscan should
however be cleaned regularly and in the event of contamination.
Do not use aggressive cleaning agents.
Do not use abrasive cleaning agents.
Static charges cause dust particles to be attracted to the front screen. You can reduce
this effect by using the anti-static plastic cleaner and the SICK lens cloth (see "12.2
How to clean the front screen:
1. Use a clean, soft brush to remove dust from the front screen.
2. Then wipe the front screen with a clean, damp cloth.
9.2
Replacing a Bulkscan
All external cable connection terminate in the system plug or the plug connections,
therefore there is no need to repeat the electrical installation when the sensor is re-
placed. The replacement sensor can simply be connected.
How to replace a Bulkscan:
Switch off the power supply for the Bulkscan.
Disconnect the connection cable from the Bulkscan.
Remove the faulty sensor.
Mount the replacement sensor (see "5 Mounting“ on page 29).
Connect the cables to the new Bulkscan or plug in the system plug in the Bulkscan.
Configure the replacement sensor by using the parameter set of the former sensor
saved on the PC (see "7 Configuration“ on page 49).
Note