MOUNTING
5
Mounting
Do not open the Bulkscan housing. If opened, any warranty claims against SICK AG are
void.
Overview of the mounting steps
1.
Placing mounting materials ready
(see "5.1 Placing mounting materials ready“ on page 29).
2.
Selecting the mounting location
(see "5.2 Selecting the mounting location“ on page 29).
3.
Mounting the sensor (see "5.3 Mounting the sensor“ on page 30).
5.1
Placing mounting materials ready
• Bulkscan, weight approx. 3.7 kg.
• Mounting kit(s) with mounting material (not included).
or
alternatively, if a mounting bracket is provided by the user:
◦
Stable mounting bracket that provides adjustable alignment of the Bulkscan on
the x and y axis.
◦
4 M6 screws for the Bulkscan, screw length dependent on the thickness of the
mounting bracket used.
• 4 M6 screws for mounting the SICK mounting bracket on the support, screw length
dependent on the thickness of the support.
• Tool set.
5.2
Selecting the mounting location
Risk of malfunction or damage due to direct sunlight on the sensor!
Direct sunlight on the sensor can cause the sensor to overheat.
Mount the sensor so that it is not exposed to direct sunlight.
Risk of malfunction due to deflection!
If the emitted laser beam is deflected by reflective surfaces, this may result in incorrect
measurements.
Avoid installing the sensor in view of glass or stainless steel surfaces as they may act
as a mirror.
• Mount the sensor so that it is protected from moisture, dirt, damage, and direct
sunlight.
•
Ensure that the entire field of view of the sensor is not restricted.
• Mount the sensor so that the status indicators can be clearly seen.
• Always mount the sensor so that there is still enough space for mounting and remov-
ing the system plug or the connections.
• Avoid excessive shock and vibration loading on the sensor.
Note