Parameter
Factory setting
Blanking size
Cross-section 200 mm
Response time of the fields
335 ms (5 scans)
Digital output holding time
335 ms (5 scans)
When selecting the response time, note that the device internal reaction time must also
be added (max. 67 ms).
To test the effects of the changed settings, change to the
Field evaluation monitor
view.
If the changed fields have been transferred to the device as described, SOPAS will
also display these in the monitor, displaying the infringed fields in yellow. If you wish to
observe another field set, it must first be activated accordingly using the digital inputs.
7.2.4
Parameter - fields
Figure 25: Device window: Parameter - fields
The user can change parameters in the right part of the window. SOPAS immediately
transfers these changes to the device (default setting).
However, evaluation fields that have been changed in size and shape must always be
manually transferred to the device using the button. All changed parameters are only
temporarily stored in the device for the time being and are not stored on the computer.
In order to optimize the dimensions of the monitoring fields:
1.
Under Field selection, for example, select field set 1.
2.
Select the field to be configured.
3.
Make the required optimizations, see description in the following sections.
Shifting field positions
1.
Button .
2.
Click on the green marking rectangle of the desired field position in the outer field.
✓
The color of the marker square changes to blue.
3.
Re-click the rectangle and drag it to the desired position, then release the push‐
button.
✓
SOPAS controls the available positioning area during shifting.