48
User Manual
"Steam sterilizer Laboklav 55 - 195
"
10.4. Changing the venting air filter
The wear of the air filter is largely dependent on the quality of the ambient air. Particularly air
contaminated with particles clogs the filter faster. Depending on the weather, this can fluctuate over
the year. High humidity shortens the life of the air filter. We recommend changing the filter after 450
cycles in winter, after 300 cycles in summer or defining a regular cycle that corresponds to the
individual frequency according to our proposal.
For devices without the S option, the filter is located at position 2 below the small cover. For devices
with a corresponding option it is placed within the filter housing at position 1.
Figure 3: Laboklav free standing unit with filter
The filter is located under the filter cover (1) on the front of the device below
the chamber door. First open the chamber door to the right.
Remove and replace the filter by prying it out.
The filter in position 2 can be removed by turning it anti-clockwise.
The new filter is screwed in clockwise accordingly.
10.5. Changing the lid seal
It is recommended to check the door seal annually and replace it if necessary. The door seal should
only be replaced properly by authorized personnel.
Improper handling / assembly can damage the seal and leakage or scalding as a result. In this case,
contact customer service!