"Steam sterilizer Laboklav 55 - 195"
User manual
45
10.1. Regular cleaning, maintenance and service activities
activity
Recommended time
notes
daily wee
kly
mon
thly
half
year
year
ly
Cleaning the surface of chamber ring
X
Cleaning chamber inside
X
Especially after over
boiling of sugar or agar
solution
Cleaning baskets
X
Cleaning trays and bottom sheets
X
Cleaning lid seal and check for
damages
X
Change lid seal if
damaged (SERVICE)
Cleaning the device outside
X
X
X
X
X
As required,
recommendation: daily
Check the safety valve(s)
X
X
SERVICE
Check the in/out connections
X
X
Change the venting air filter
X
X
X
After message by device
Function test for the valves
X
SERVICE
Cleaning the tank
X
SERVICE
Functional test of vacuum pump
X
SERVICE
Check program parameters
X
X
Check for lid / door adjustment
X
SERVICE
Check steam generator
X
SERVICE
Electrical test
X
SERVICE
Attention please! Opening the unit is allowed for authorized and trained service stuff only!
If other intervals are prescribed at the place of use according to national law, then these
inspection periods and periods must be observed!
The vessel has to undergo regular inspections and tests
according to the current laws and regulations. We advice you
to have an inspection of the vessel jacket regularly to check
any corrosion. Such inspection has to be undertaken by an
authorized person of our company.
Attention: Please pay attention when the lowest stress cycle
has reached the half amount you have to inspect the
pressure vessel. The inspection has to be done in Germany
due to BetrSichV not later than 10 years. Out of Germany
local regulations have to be obeyed.