4.4. First commissioning
Follow these steps one after the other to get the device started and running for the first time:
4.4.1. Setting up the device
You can set up the device yourself and put it into operation. It is essential to observe the following
notes on setting up, installation and commissioning in this operating manual! It is recommended that
the autoclave is set up, installed and commissioned by specialist personnel who has been authorized
by SHP Steriltechnik AG for the first commissioning.
The sterilizer should be set up on a flat surface. When using the sterilizer as a table-top unit, the
optionally available base frame should be used.
Please leave 10 cm of space between the rear and
side walls of the device and the surroundings. The room should be well ventilated, the heat
emission of the device to the surroundings should not be impeded.
The device should be
aligned with the help of the adjustable feet so that the condensate can always drain into the chamber
outlet. The snorkel should be set up accordingly.
Do not place the device immediately under a smoke detector!
Otherwise, depending on the type of the detector, it may
trigger a false alarm due to rising water steam. The device
should therefore always be set up at least 3 meters away
from the smoke detector. If necessary, the smoke detector
should be replaced by a temperature-dependent fire alarm.
4.4.2. Switching on
After regular installation and connecting to media / power supply the device is ready for use. Standing
in front of the device you will find the main switch at the right side of the key pad. Switching on the
main switch the display is switched on and shows the software version and the SHP Logo short time.
When switched on the device is ready for use.
Do not manipulate the device! If the device is not going on
please check the main breakers of the unit and of power
supply. If no reason can be found please inform a trained
service stuff to check the device internally.
After switching on and going into a program (without P12 –
vacuum test) the device is preheating the steam generator
when lid is closed!