4.4.9. Changing program parameters
To change program parameters press the
Enter button (7)
. You reach the code menu.
Give the
code 2000
to the menu and confirm with Enter button again.
To navigate through the menu use the cursor buttons.
If the code was correct you reach the main menu. Depending on the access level defined by the code
the main menu shows different submenus.
In the picture you find the maximum main menu, by giving the code 2000 you find the program
parameter point only.
Enter the point program parameters. Inside the submenu go to the program you want to change.
Inside this submenu you can choose the program phase that
should be changed directly.
Change Program parameters only if the result gives real
advantage! The preprogrammed sterilization cycles are
validated for empty chamber and full loaded chamber. For
the normal use the 10 preprogrammed cycles should be
enough.
When all programs are configured you should run and test the program with empty chamber and full
loaded chamber to be sure the program parameters do not make problems in normal cycle run. If
there are doubts, you should make a full validation of the program.