"Steam sterilizer Laboklav 55 - 195
"
Liquids may only be sterilized if a thermal barrier is active. This happens automatically in all liquid
programs. Therefore only use the corresponding liquid programs for liquids! In order to ensure the
correct functioning of the thermolock and the corresponding program, it is necessary that the
reference sensor is placed in a vessel containing as much water as its largest individual volume in the
chamber!
Ist der Referenzfühler nicht richtig platziert oder befindet sich
gar außerhalb des Referenzgefäßes, kann es beim Öffnen des
Gerätes zu Siedeverzug und damit zu Verletzungen durch
Verbrühung kommen. Außerdem wird die Sterilisierzeit zu
früh ausgelöst!
Attention please while using hot liquids! While contact with
hot liquids with temperature of more than 60°C it can burn
the skin!
Attention please while handling closed waste bags! If the unit
is not equipped with vacuum the waste needs to be opened
while loaded into the chamber. While opening the waste bag
bio aerosols will come free and may infect the operator!
Operate the device in the right protection clothes only!
Protect your skin, your face especially eyes, nose, mouth!
7.3. Removal of condensate
To remove the condensate the unit is equipped with an outlet to connect to a drainage system. If the
unit is connected with cold water supply the program is automatically protecting the condensate drain
against over heating by direct draining with hot steam.
If there is no possibility of disposal, we offer you the option of purchasing appropriate condensate
collection containers. Ask your dealer or contact us directly.