34
Insulated power
output
(P24 option)
If this option is ordered, the A2 function will be disabled.
This option cannot be used with the Insulated power output (P5 option) together,
or cannot be used with Power for 2-wire transmitter (DSB option) together.
Output voltage
24 3 V DC (When load current is 30 mA)
Ripple voltage
Within 200 mV DC (When load current is 30 mA)
Max load current
30 mA DC
Insulated power
output
(P5 option)
If this option is ordered, the A2 function will be disabled.
This option cannot be used with the Insulated power output (P24 option) together,
or cannot be used with Power for 2-wire transmitter (DSB option) together.
Output voltage
5 0.5 V DC (When load current is 30 mA)
Ripple voltage
Within 200 mV DC (When load current is 30 mA)
Max load current
30 mA DC
Power for
2-wire
transmitter
(DSB option)
If this option is ordered, only 4 to 20 mA DC input (Built-in 50
shunt resistor)
can be used.
This option cannot be used with the Insulated power output (P24 option or P5
option) together.
Output voltage
24 3 V DC (When load current is 30 mA)
Ripple voltage
Within 200 mV DC (When load current is 30 mA)
Max load current
30 mA DC
Transmission
output 2
(T
□
2 option)
This option cannot be used with Alarm 4 output (A4 option) together.
Resolution
12000
Output accuracy
Within
0.3% of transmission output span
Response time
400 ms + Input sampling period (0%
→
90%)
Option Code
Transmission Output Type
TA2 (4-20)
Direct
current
4 to 20 mA DC (Load resistance: Max 550
)
TA2 (0-20)
0 to 20 mA DC (Load resistance: Max 550
)
TV2 (0-1)
DC
voltage
0 to 1 V DC (Load resistance: Minimum 100 k )
TV2 (0-5)
0 to 5 V DC (Load resistance: Minimum 500 k )
TV2 (1-5)
1 to 5 V DC (Load resistance: Minimum 500 k )
TV2 (0-10)
0 to 10 V DC (Load resistance: Minimum 1 M )
User specified
Transmission
output
(TA, TV option)
Standard Transmission output can be changed to the following ‘User specified
Transmission output’.
Option Code
Transmission Output Type
TA (0-20)
Direct
current
0 to 20 mA DC (Load resistance: Max 550
)
TV (0-1)
DC
voltage
0 to 1 V DC (Load resistance: Minimum 100 k )
TV (0-5)
0 to 5 V DC (Load resistance: Minimum 500 k )
TV (1-5)
1 to 5 V DC (Load resistance: Minimum 500 k )
TV (0-10)
0 to 10 V DC (Load resistance: Minimum 1 M )
Color Black
(BK option)
Panel: Dark gray
Case: Black
Terminal cover
(TC option)
Electrical shock protection terminal cover