8. FUNCTIONAL DIAGRAM
VM-5S
-60-
10
Level adjustment
Converts the DC input voltage to the level required by the A/D converter.
11
A/D
10-bit A/D converter for conversion of analog to digital signals.
12
Counter (Enabled for VM-5C, 5 S, and 5R)
Counts the pulse input from the trigger card.
13
CPU
16-bit CPU used.
14
PWM
Performs pulse width modulation in proportion to the measured value.
15
SIO
Handles internal communication with the VM-5P1,2 Communication/Phase Marker Unit.
16
PIO
Inputs signals for key operations and remote contact input, and outputs signals for driving the
LCD cards and alarm relays.
17
Integrator
Integrates the PWM (pulse width modulation) signal, and converts it to a DC voltage
proportional to the measured value.
18
Output amplifier
Converts the output to the 1 to 5VDC or 4 to 20mA signal required for the recorder, etc.
19
Filter card (option)
Available as option, if a still sharper filter performance is required.
20
Trigger card (Enabled for VM-5C, 5 S, and 5R)
Converts the input signal to the pulse waveform.
21
Isolated output card, voltage conversion card (option)
(0 to -10VDC, 0 to +10VDC, 0 to -5VDC, 0 to +5VDC)
• Isolated output card :
This isolated output card is used for isolation from other circuits.
• Recorder option output card :
A recorder option output card for applications where an other than standard recorder output
is required.