CAA-100 Series User’s manual
Shineway Technologies, Inc.
-
37-
In the measurement of interface, click on the button "cal mode"
The OSL calibration procedure is as follows,
Set the frequency parameter
Set up calibration parameters (optional)
Calibrate the first load
Connect the load to the RF ports, click on the corresponding button.
The system will pop up a small window, click on the “OK”.
System start calibration.
Once calibration is completed the screen will display the curve.
Calibrate other loads (Repeat the above steps)
If all calibration of three loads completed, the green "cal on" will shown at the lower left corner of the
screen.
Flexcal mechanical calibration is as the following step:
Set the frequency parameter and measurement point
Select the Flexcal calibration
Click open / short / load under calibration interface and select the corresponding calibrator to calibrate
If all calibration of three loads completed, the green "cal on" will shown at the lower left corner of the
screen.
Flexcal electronic calibration is as the following step:
Set the frequency parameter and measurement point
Select the Flexcal calibration
Connect the electronic calibrator and USB cable to the unit
Click the ECAL coin, and it will be automatic to calibrate
If all calibration of three loads completed, the green "cal on" will shown at the lower left corner of the
screen.
Users can directly calibrate at the RF port, also can calibrate with a high performance cable connected to RF
port. If the former, usually, the calibration curve of 50 ohm load will be significantly lower than the other two.
If calibration is completed, the user can start measurement .
The user can also select the file menu to save the calibration data in the local or external disk.
After the system boot, the default calibration data will be recent one.
The user clicks on the "file" menu, click "import", can recall previous calibration data. If the environment is
similar, the previous calibration data can be used directly.
For mechanical calibration, users need to re calibration if frequency changed or measurement points
increased.
For Flexcal calibration, users do not need to re calibration if frequency changed or measurement points
increased.
Содержание CAA-100 Series
Страница 1: ...CAA 100 Series Cable and Antenna Analyzer User s Manual...
Страница 25: ...CAA 100 Series User s manual Shineway Technologies Inc 19 Image preview interface File import interface...
Страница 33: ...CAA 100 Series User s manual Shineway Technologies Inc 27 Power mode setting interface...
Страница 48: ...CAA 100 Series User s manual Shineway Technologies Inc 42 THANK YOU FOR CHOOSING SHINEWAY TECHNOLOGIES...