CAA-100 Series User’s manual
Shineway Technologies, Inc.
-
23-
the instruction and connect open / short/50 Ohm load respectively to the RF port and do calibration. Finishing the
mechanical calibration, if users change frequency or the measurement point, this calibration will not effect and
users should recalibrate.
Electronic calibration is done with electronic calibrator, Flexcal calibration. Connect electronic calibrator and
data cable to the unit, click the Ecal coin, it will do the electronic calibration. Electronic calibration is default in the
whole frequency, and the measurement point is 551; Finished the electronic calibration, it will be effect whether
users change the frequency range and measurement point or not.
When users select the Flexcal, the calibration is default the whole frequency range calibration and the
measurement range is 551. When calibration is finished, changing the frequency range or measurement point will have
no effect on the calibration status.
Mechanical calibration can select the Flexcal or not. If select the Flexcal, calibration state will not be effect by
changing the frequency range or measurement point; if not, when users change the frequency range or measurement
point, users must do recalibration.
Electronic calibration is default to select the Flexcal.
Calibration interface
The calibration interface is similar to measurement interface. On the left-top of screen including frequency
information, On the left-bottom of screen including calibration information.
The user can save current calibration data to the local or external disk, or Import history calibration data to
Содержание CAA-100 Series
Страница 1: ...CAA 100 Series Cable and Antenna Analyzer User s Manual...
Страница 25: ...CAA 100 Series User s manual Shineway Technologies Inc 19 Image preview interface File import interface...
Страница 33: ...CAA 100 Series User s manual Shineway Technologies Inc 27 Power mode setting interface...
Страница 48: ...CAA 100 Series User s manual Shineway Technologies Inc 42 THANK YOU FOR CHOOSING SHINEWAY TECHNOLOGIES...