CAA-100 Series User’s manual
Shineway Technologies, Inc.
-
24-
CAA-100 Series.
Before calibration, the user need to set the correct parameters, Mainly refer to the frequency and
measurement points. The completion calibration, the corresponding curve will display on the screen. If calibration
is valid, the status on the left-bottom will become “cal On”. If calibration is invalid or unfinished, the status will be
“cal off”.
After the calibration, the user can return directly to the measurement interface for measurement.
System support different calibration kit only if user know the exact electrical length of open/short load.
Users can input this parameters in the e parameter menu.
Is
the
electrical
length of open and
short cal kit same?
Is the exact electrical
length of open and short
cal kit known?
Return
Loss/VSWR/cable Loss
measurement
Phase/Impedenc-e
measurement
Yes
unknown
support
Do not support
Yes
know
support
Support
know
support
support
The return loss of 50 ohms cal kit need to be > 42.
The VSWR of open/short cal kit need to be >100.
If the user does not focus on phase and impedance characteristics, cal kit from other manufacturers can be
directly used (usually short and open cal kit have same electrical length).
Calibration setting menu
Note: the electric length is measured at 1GHz frequency, the unit is degree.
Содержание CAA-100 Series
Страница 1: ...CAA 100 Series Cable and Antenna Analyzer User s Manual...
Страница 25: ...CAA 100 Series User s manual Shineway Technologies Inc 19 Image preview interface File import interface...
Страница 33: ...CAA 100 Series User s manual Shineway Technologies Inc 27 Power mode setting interface...
Страница 48: ...CAA 100 Series User s manual Shineway Technologies Inc 42 THANK YOU FOR CHOOSING SHINEWAY TECHNOLOGIES...