CAA-100 Series User’s manual
Shineway Technologies, Inc.
-
34-
Select pre-defined Signal standard
Manual input frequency parameter
Users can click on edit box to edit the parameter of frequency , system will pop up the soft keyboard. The user
also can directly press the digital keys of hard keyboard .
The users need to select measurement points . System supports 4 options: 137,275,551,1103.
If the measurement point is set to 1103, compared with the set to 551, it will take about 2 times longer
measurement time. And so on.
The minimum frequency interval is 1MHz.
The start frequency range is 25~3999MHz;The stop frequency range is 26~4000MHz.
The user can modify the pre-defined signal standard through PC software .
4.2 Set DTF parameter
The user needs to properly set the DTF parameter before DTF testing. In the following ways users can enter
the DTF setting interface.
Enter the "DTF-return loss" or "DTF-VSWR" measurement interface, select "parameter" menu.
DTF setting interface
The user can complete the following functions at the DTF parameter setting interface
Enter frequency information and measure points
Enter the distance information
Enter the cable parameters
Select the window function
Set unit (support metric and Imperial)
Содержание CAA-100 Series
Страница 1: ...CAA 100 Series Cable and Antenna Analyzer User s Manual...
Страница 25: ...CAA 100 Series User s manual Shineway Technologies Inc 19 Image preview interface File import interface...
Страница 33: ...CAA 100 Series User s manual Shineway Technologies Inc 27 Power mode setting interface...
Страница 48: ...CAA 100 Series User s manual Shineway Technologies Inc 42 THANK YOU FOR CHOOSING SHINEWAY TECHNOLOGIES...