
49 |
P a g e
MAINTENANCE
Continued next page
Temperature Calibration Continued
3
Place the display in temperature calibration mode.
a.
Press and hold both the
Up
and
Down
temperature
arrow buttons simultaneously for approximately 5
seconds.
b.
Release the buttons when the temperature display
shows the letters “C O”. The display will begin
flashing the
current temperature display value
.
Note:
The display will automatically exit calibration mode after
5 seconds of inactivity, with the last shown temperature
display value saved.
4
Use the
Up
or
Down
arrows to adjust the current
display temperature value until it matches the
reference device temperature reading.
Reference Device
5
After matching the display to the reference device,
wait 5 seconds.
•
The temperature display will cease flashing
and store the corrected chamber display value.
•
The incubator will now begin heating or
passively cooling in order to reach the setpoint
with the corrected display value.
Cooling to Setpoint
6
Allow the incubator to operate for at least 1 hour
undisturbed to stabilize
after the incubator has
achieved the corrected temperature setpoint.
•
Failure to wait until the incubator is fully
stabilized will result in an inaccurate reading.
Setpoint Achieved
Содержание SCO10A
Страница 1: ...AIR JACKETED CO2 INCUBATOR 110 120 Volts Installation Operation Manual SCO10A SCO5A...
Страница 6: ...6 P a g e TABLE OF CONTENTS...
Страница 24: ...24 P a g e INSTALLATION...
Страница 26: ...26 P a g e SYMBOLS...
Страница 59: ...59 P a g e PARTS...