40 |
P a g e
OPERATION
L
OADING
S
AMPLES
The manufacturer strongly recommends waiting at least
8 hours
after putting the incubator in
operation before loading samples in the chamber. This safeguards against temperature instability.
•
Samples should be placed at least 1 inch (25 mm) away from the chamber walls.
•
Samples or sample stacks should be spaced 1 inch (25 mm) apart.
•
Proper spacing allows for maximum air circulation and a higher degree of temperature
uniformity.
•
Proper spacing also decreases the chance of condensate forming in the incubator when
operating with a large number of samples in the chamber.
A
CCESSORY
C
OMPATIBILITY
•
Verify that any powered accessory equipment used inside the chamber can safely and
effectively operate within your selected temperature, humidity, and CO
2
concentration
range.
•
Powered equipment, such as stirrers or shakers, can generate heat sufficient to disrupt the
thermal uniformity and stability of the chamber.