USER MAINTENANCE 

Warning

: Disconnect the unit from its power supply prior to performing maintenance or services. 

Avertissement

: Débranchez cet appareil de son alimentation électrique avant d'effectuer la maintenance 

ou les services.

 

 

C

LEANING AND 

D

ISINFECTING

 

If a hazardous material or substance has spilled in the unit chamber, immediately initiate your site 

Hazardous Material Spill Containment protocol. Contact your local Site Safety Officer and follow 

instructions per the site policy and procedures. 

 

Periodic cleaning and disinfection are required. 

 

Do not use spray-on cleaners or disinfectants. These can leak through openings and coat 

electrical components. 

 

Consult with the manufacturer or their agent if you have any doubts about the 

compatibility of decontamination or cleaning agents with the parts of the equipment or 

with the material contained in it. 

 

Do not use cleaners or disinfectants that contain solvents capable of harming paint 

coatings or stainless-steel surfaces. 

Do not use chlorine-based bleaches or abrasives; 

these will damage the chamber liner.

 

 

Do not clean or disinfect the ring-style chamber HEPA filter.

 Replace the filter if 

discolored or if you believe it has been contaminated. 

 

Warning

: Exercise caution if cleaning the unit with alcohol or flammable cleaners. Always allow the 

unit to cool down to room temperature prior to cleaning and make sure all cleaning agents have 

evaporated or otherwise been completely removed prior to putting the unit back into service. 

Avertissement:

 

Soyez prudent lorsque vous nettoyez l'appareil avec de l'alcool ou des produits de 

nettoyage inflammables. Laissez toujours refroidir l'appareil à la température ambiante avant le 

nettoyage et assurez-vous que tous les produits de nettoyage se sont évaporés ou ont été 

complètement enlevés avant de remettre l'appareil en service.

 

 

Cleaning 

 

Disconnect the unit from its power supply. 

 

Remove all removable interior components such as shelving and accessories. 

 

Clean the unit with a mild soap and water solution, including all corners. 

o

 

Do not

 

use abrasive cleaners

, these will damage metal surfaces. 

o

 

Do not use deionized water to rinse or clean with.

 

o

 

Take special care when cleaning around the temperature sensor probes in the 

chamber to prevent damage. Do not clean the probes. 

 

Rinse with distilled water and wipe dry with a soft cloth. 

Содержание SCO10A

Страница 1: ...AIR JACKETED CO2 INCUBATOR 110 120 Volts Installation Operation Manual SCO10A SCO5A...

Страница 2: ...Advertencia Este producto contiene sustancias qu micas incluido el triglicidil isocianurato que el estado de California sabe que causa c ncer as como defectos de nacimiento u otros da os reproductivos...

Страница 3: ...rs for professional industrial or educational use where the preparation or testing of materials is done at an ambient air pressure range of 22 14 31 3 inHg 75 106 kPa and no flammable volatile or comb...

Страница 4: ...ion 21 Connect to the CO2 Supply 22 Access Port Stopper 23 GRAPHIC SYMBOLS 25 CONTROL PANEL OVERVIEW 27 OPERATION 29 Theory of Operation 29 Put the Incubator into Operation 31 Humidify the Incubator 3...

Страница 5: ...5 P a g e UNIT SPECIFICATIONS 55 Weight 55 Dimensions 55 Max Shelves For Unit 56 Capacity 56 CO2 56 Temperature 56 Power 56 PARTS LIST 57 Accessories 58 Ordering 58...

Страница 6: ...6 P a g e TABLE OF CONTENTS...

Страница 7: ...w up a site specific standard operating procedure SOP covering each application and associated safety guidelines This SOP must be written and available to all end users in a language they understand I...

Страница 8: ...ufacturing INC P O Box 627 Cornelius OR 97113 USA MANUFACTURING WARRANTY For information on your warranty and online warranty registration please visit sheldonmanufacturing com warranty ENGINEERING IM...

Страница 9: ...ng contaminants into the chamber may damage the incubator and is not covered by the manufacturing defect warranty Supply Source and Pressure The incubator requires 15 20 psi of CO2 gas pressure at the...

Страница 10: ...est CO2 accuracy use a digital gas analyzer with sample tubing connecting to the incubator CO2 sample port The barbed adapter gas fitting on the sample port connects to a 3 16 inch 4 76 mm inner diame...

Страница 11: ...er service that delivered the unit 3 If the carton is not damaged open the carton and remove the contents 4 Inspect the unit for signs of damage Use the orientation images in this chapter as reference...

Страница 12: ...rt Fuse in Power Inlet USB Style Port Control Panel CO2 Sample Port Side of Unit Chamber Gasket Ceiling Air Duct Chamber HEPA Filter Door Latch Humidification Pan Viewing Door Glass Chamber Door Chamb...

Страница 13: ...13 P a g e RECEIVING Top Incubator SCO10A Mounting Bracket and SCO10 A Data Plate Bottom Incubator See the SCO5A depiction on the previous page for details on the individual incubators in the stack...

Страница 14: ...this information to provide accurate help during support calls and emails SCO5A o The data plate is located on the inside of the incubation chamber door in the top right corner SCO10A o The data plat...

Страница 15: ...able electrical outlet and power supply is present page 17 Install the incubator in a suitable location Review the lifting and handling instructions page 18 Install the unit leveling feet page 18 Inst...

Страница 16: ...unit consider all environmental conditions that can adversely impact its temperature performance These include Proximity to ovens autoclaves and any device that produces significant radiant heat Heati...

Страница 17: ...n 10 from the data plate rating Use a separate circuit to prevent loss of the unit due to overloading or circuit failure The recommended wall circuit breakers for these units are 15 amps The wall powe...

Страница 18: ...r stabilization Restrain the unit completely while lifting or transporting so it cannot tip Remove all moving parts such as shelves and trays and lock doors in the closed position during transfers to...

Страница 19: ...Ohm cm or a conductivity range of 20 0 uS cm to 1 0 uS cm for cleaning applications INSTALLATION CLEAN AND DISINFECT The manufacturer recommends cleaning the shelving ceiling air duct and chamber prio...

Страница 20: ...eps to install the filter and ceiling air duct Snap the chamber HEPA filter to the mounting collar on the chamber ceiling air duct It may be necessary to tilt the filter to one side or the other Snap...

Страница 21: ...Align the keyhole slot of the standard with the mounting peg on the side of the chamber wall Mount the shelf standard Install the shelf slides Insert the shelf slide into the shelf standard using a ro...

Страница 22: ...upply source to each incubator The SCO10A requires a CO2 supply for each chamber Do not initiate a flow of CO2 to the incubator at this time End of Procedure CO2 to Chamber IN Connects to 3 16 4 76 mm...

Страница 23: ...nside the chamber to obtain the best temperature uniformity and prevent condensation from forming inside the port o Do not install the stopper on the outside of the incubator Wires for thermocouples a...

Страница 24: ...24 P a g e INSTALLATION...

Страница 25: ...ichage de la temp rature Over Temperature Limit system Thermostat temp rature limite contr le haute AC Power Rep re le courant alternatif I ON O OFF I indique que l interrupteur est en position marche...

Страница 26: ...26 P a g e SYMBOLS...

Страница 27: ...ing to the incubation chamber by rerouting power away from the heating elements Set Temperature Display and Controls Shows the current chamber temperature The Up and Down arrow buttons are used to acc...

Страница 28: ...to access the CO2 Setpoint SP or Calibration Offset C O display modes and input the CO2 setpoint or calibration adjustment value Red LED alarm indicators marked High and Low illuminate whenever the CO...

Страница 29: ...l room it will increase the length of heating pulses to compensate Likewise when operating in a warm room the incubator uses shorter pulses If the ambient temperature conditions change significantly t...

Страница 30: ...chamber The system operates independently of the main microprocessor temperature controller and routes power away from the incubator heating elements if the chamber temperature exceeds the OTL tempera...

Страница 31: ...cord that came with the unit to the power inlet receptacle on the side of the incubator Plug the power cord into the workspace electrical supply outlet 3 Turn on the incubator Place the incubator Powe...

Страница 32: ...t for at least 8 hours for example overnight with chamber doors closed prior to Setting the Over Temperature Limit next step Loading samples This helps ensure a stable temperature environment 9 Perfor...

Страница 33: ...o slow the growth of microbiological populations in the humidification water supply Never use deionized water to humidify the incubator The pan must be placed on the chamber floor o The floor is heate...

Страница 34: ...temperature setpoint adjustment mode The display will briefly flash the letters SP then show the flashing adjustable temperature setpoint Note The display will automatically exit the adjustment mode a...

Страница 35: ...r the duration of the current temperature deviation The visual alarm indicator will remain illuminated Any new deviation of 1 C or greater will reactivate the audible alarm AUTOMATIC DOOR CUTOFFS Gas...

Страница 36: ...ply CO2 to the regulator Set the cylinder regulator flow setting Set the regulator to deliver 15 20 Pounds per Square Inch psi to the incubator CO2 in Port Never exceed 20 psi Building Supply System I...

Страница 37: ...buttons to activate the CO2 setpoint adjustment mode The display will briefly flash the letters SP then show the flashing adjustable CO2 setpoint Note The display will automatically exit the adjustmen...

Страница 38: ...off The audible alarm will remain muted for the duration of the current CO2 deviation The visual alarm indicator will remain illuminated Any new deviation of 1 or greater will reactivate the audible a...

Страница 39: ...activating check for ambient sources of heat or cold that may be adversely impacting the unit temperature stability If you find no sources of external or internal temperature fluctuations contact Cust...

Страница 40: ...r spacing allows for maximum air circulation and a higher degree of temperature uniformity Proper spacing also decreases the chance of condensate forming in the incubator when operating with a large n...

Страница 41: ...Outputs The temperature and CO2 levels are also outputted as analog signals by an internal 4 20 milliamp board The outputs are transmitted through two jack ports on the back of each incubator unit Th...

Страница 42: ...y in the incubation chamber is already near the dew point barometric fluctuations may push it across the dew point threshold Note Thin air at higher altitudes holds less humidity than the denser air f...

Страница 43: ...iner Do not clean or disinfect the ring style chamber HEPA filter Replace the filter if discolored or if you believe it has been contaminated Warning Exercise caution if cleaning the unit with alcohol...

Страница 44: ...rom evaporating disinfecting agents can inhibit growth or cause metabolic symptoms in microbiological sample populations Make sure that chlorines quaternary ammonias or any other overtly volatile disi...

Страница 45: ...ss permanent kinking or other signs of damage are present ELECTRICAL COMPONENTS Electrical components do not require maintenance If the incubator fails to operate as specified please contact your dist...

Страница 46: ...er the front of the duct This safeguards against accidentally striking the blower fan and sensor heads with the duct Pull the duct and attached HEPA filter out of the incubation chamber Replacing the...

Страница 47: ...must be humidified for at least 24 hours in order to conduct an accurate calibration A Suggested Calibration Set Up CO2 Calibration A CO2 calibration may be performed during the temperature calibrati...

Страница 48: ...Calibration 1 Once the incubator temperature has stabilized compare the reference device and incubator temperature display readings If the readings are the same or the difference between the two fall...

Страница 49: ...saved 4 Use the Up or Down arrows to adjust the current display temperature value until it matches the reference device temperature reading Reference Device 5 After matching the display to the referen...

Страница 50: ...ubator is now calibrated for temperature OR See the next step if the readings fail to match or fall outside of your protocol range Reference Device 8 If the difference still falls outside the acceptab...

Страница 51: ...ten as required by your laboratory or production protocol or regulatory compliance schedule Always calibrate to the standards and use the calibration setup required by your industry requirements or la...

Страница 52: ...ested CO2 Calibration 1 Once the incubator CO2 concentration has stabilized compare the reference device and the CO2 display readings If the readings are the same or the difference between the two fal...

Страница 53: ...Down arrow buttons to adjust the current display CO2 value until it matches the reference device concentration reading Reference Device 5 After matching the display to the reference device wait 5 seco...

Страница 54: ...or is now calibrated for CO2 OR See the next step if the readings fail to match or fall outside of your protocol range Reference Device 8 If the difference still falls outside the acceptable range of...

Страница 55: ...inner chamber All indications are average values typical for units produced in the series We reserve the right to alter technical specifications at all times WEIGHT Model Shipping Unit SCO5A 238 lb 10...

Страница 56: ...0 142 0 SCO10A 5 0 142 0 Each Chamber CO2 Model Range Accuracy Recovery Time All 0 20 0 1 Less than 5 minutes TEMPERATURE Model Range Uniformity Stability All Ambient 5 C to 60 C 0 25 C at 37 C 0 1 C...

Страница 57: ...cation Pan 995 00015 Chamber HEPA Filter Cap 6500506 Leveling Foot 2700512 Ceiling Air Duct Chamber 5121527 Power Cord 115 volt 15 Amp 9ft 5 in 2 86m NEMA 5 15P 1800510 CO2 Gas Tubing Kit with In Line...

Страница 58: ...ssembly described above PN 5820504 Individual Shelf PN 5820505 Pair of Slides ORDERING Accessories and replacement parts can be ordered online at parts sheldonmfg com If the required item is not liste...

Страница 59: ...59 P a g e PARTS...

Страница 60: ...P O Box 627 Cornelius OR 97113 USA support sheldonmfg com sheldonmanufacturing com 1 800 322 4897 503 640 3000 FAX 503 640 1366...

Отзывы: