8.9 Panel
The panel option allows the user to restrict the function of the front panel Hot
Keys. If the option is selected, the Hot Keys are restricted.
While panel is displayed press the to enter the sub menu. The main display
now displays either
on
or
oFF
. Use the key to select the required status.
Press to confirm selection.
8.10 Reset
The
rESEt
sub menu allows the user to reset the instrument back to the default
settings. If the option is protected the user will be required to enter the password.
9 AcuDew - AutoCal
9.1 Benefit of AutoCal
AutoCal allows the user to ensure accuracy to the laboratory calibration by
checking the span of the transmitter and correcting for any deviation. It should be
operated periodically, every 2 to 3 months, or when verification of the AcuDew
Transmitter is required.
35