Units Hot Key
The ‘Units’ hot key allows the user to view and alter the displayed units.
To review the moisture in alternative units, press the (Units) key for longer
than 3 seconds. The main display will then show the message
UnitS
. Press the
key to enter the routine. Pressing the key displays the current moisture level in
each of the moisture units in turn. Press the key at any time sets the currently
displayed units as the AcuTrak active unit.
For example: To change from a ‘dewpoint °C’ to a ‘dewpoint °F’, whilst
UnitS
is
displayed press the key to enter the subroutine.
The secondary display now shows the unit type flashing. Use the key to scroll
through to the next unit type is displayed e.g. °F
Press the key to select the new unit.
Note:
Pressing the key at any time escapes the hot key function and returns to
the measurement display without saving any changes.
Alarms Hot Keys
The two alarm hot keys ‘AL 1’ and ‘AL 2’ allow the user to review and set the alarm
trigger points.
To review the alarm trip point press the (AL1) or (AL2) key momentarily.
The primary display will show the set trip point for 1 second before reverting back
to the moisture reading.
To alter the trip point press and hold the (AL1) or (AL2) keys for longer than
3 seconds. The main display will then show the message
AL1
or
AL2
. Press the
key to enter the routine. Use the and keys to display the required alarm
trip point.
Press the key to set the trip point.
ESC
27