- 5 - 

 

SI 460A DIN SIZE WEIGHING INDICATOR 

3. SPECIFICATION 

3-1 Specification

 

Content 

Specification 

Performance 

External Resolution 

1/20,000 

Internal Resolution 

1/2,000,000 (±1,000,000) 

Input Sensitivity 

Min 0.1µV/V 

Max Signal Input Voltage 

3.0mV/V 

Load cell Excitation 

DC +5V 

A/D Conversion Method 

Sigma-Delta 

Decimal Point 

0, 0.0, 0.00, 0.000 

Drift 

Offset 

10PPM/℃ 

Span 

10PPM/℃ 

Linearity 

0.001% of Full Scale 

Analogue Sampling(sec) 

Max 60times / sec 

Environment 

Operating Temperature Range 

-10℃ ~ +40℃ [14℉ ~ 104℉] 

Operation Humidity Range 

40% ~ 85% RH, Non-condensing 

Function 

Calibration Mode 

Test Weight Calibration Mode 

Simulation Calibration Mode 

Display 

6 digit, 15.24mm(0.6inch) 

Red Color FND 

Key Pad 

5EA Standard Key pad 

Additional Digital Input 

4pcs external input key addable 

Communi- 

cation 

Standard Serial Interface 

(RS-422/485) 

Data Transference 

Command Mode 

Serial Printer Mode 

Modbus (RTU) 

Option Serial Interface 

(RS-232C) 

Data Transference 

Command Mode 

Serial Printer Mode 

Analog Output (Option) 

0~10V, 4~20mA (Selectable) 

Power 

DC : 24V,    Power Consumption 12W 

AC (Option) : 110~220V, Power Consumption 12W 

Size 

96mm(W)ⅹ96mm(H)ⅹ116mm(D)

 

Weight : 500g (DC) 

                700g (AC.) 

Содержание SI 460A

Страница 1: ...1 Digital Weighing Indicator SI 460A User Manual 2014 06 13...

Страница 2: ...2 Installation Components 9 Page 9 Page 4 3 Load Cell Installation 10 Page 5 Set up 11 Page 5 1 TEST Weight Calibration Mode 11 Page 5 2 Simulating Calibration Mode 13 Page 5 3 F FUNCTION Setting 20...

Страница 3: ...f wrongly used Copy Rights 1 All Right and Authority for this Manual is belonged to SEWHA CNM CO LTD 2 Any kinds of copy or distribution without permission of SEWHA CNM CO LTD will be prohibited 3 Thi...

Страница 4: ...ce where high voltage or heavy electric noise condition 4 When you connect with other devices please turn off the power of item 5 Avoid from water damage 6 For the improvement of function or performan...

Страница 5: ...rature Range 10 40 14 104 Operation Humidity Range 40 85 RH Non condensing Function Calibration Mode Test Weight Calibration Mode Simulation Calibration Mode Display 6 digit 15 24mm 0 6inch Red Color...

Страница 6: ...ction is set ON HOLD When HOLD function is set ON TxD When the Indicator transmits Serial communication data Print data Lamp is ON RxD When the Indicator receives Serial communication data Lamp is ON...

Страница 7: ...ction setting Save the value go to next step 3 Test Mode 1 Standard serial interface test mode 4 Set up Mode Set point setting Mode Setup Mode It is a mode can SET UP the calibration Function of SI 46...

Страница 8: ...TxD TxD RxD RxD RS 485 Standard Not used Not used RxD RxD 4 Load cell Input EXC EXC SIG SIG SHIELD 5 Analogue Output terminal Option 4 20mA Factory Default 0 10V Selectable 6 Option Serial Interface...

Страница 9: ...9 SI 460A DIN SIZE WEIGHING INDICATOR 4 INSTALLATION 4 1 External Dimension Cutting Size 4 2 Installation Components SI 460A User Manual External Dimension Unit mm Cutting Size Unit mm...

Страница 10: ...terminal correctly Do not weld near the load cells Indicators or other devices Load Cell Installation 1 You can connect Max 8pcs of same capacity Load cells at once 350 2 You have to make horizontal...

Страница 11: ...mode is ON Shortcut F function Mode Press key 4 times Test Mode 1 Analog Variation Value Press key 4 times Display Press key 4 times Key Press key 4 times Analog Output Press key 4 times Test Mode 2...

Страница 12: ...ration mode TARE HOLD PRINT will be reset Before processing calibration please warm up the indicator during 15 min to guarantee more preciseness Calibration Key Operation CANCLE BACK Move to left Move...

Страница 13: ...INDICATOR 5 2 3 Setting Capacity of weighing Scale After displaying CAPA input max capacity with keys Press key to save move to next step If you want to set Max capacity as 1 000kg and the division is...

Страница 14: ...tep Max decimal point will be 0 001 and digit can be selected among 1 2 5 10 20 50 Digit and decimal point must be fulfilled under the below condition division value Max capacity value cannot be over...

Страница 15: ...eness DEAD weight calculation CAL00 CAL09 will be operated twice In this step if there is unstable condition such as some forces or Vibration on the scale part ErrorA will be displayed and DEAD value...

Страница 16: ...part and press key Over than 1 10 000 resolution setting To guarantee the preciseness Span calculation will be operated twice After calculation span value will be displayed on the display Then press...

Страница 17: ...out any TEST weight This calibration mode uses Load cells max capacity and Max Output Rate mV so the weight adjustment degree might be less than Test weight Calibration The guaranteed resolution of th...

Страница 18: ...or Test Report Input the Max Capacity of Load cell And press key In case of multiple pieces of load cells are installed Please make sum of each load cell s capacity and make setting with Max Capacity...

Страница 19: ...oint will be changed Please stop at the optimal position And select Division optimal division with key Finally press key to save and move to next step 5 3 4 Measuring the DEAD Weight of Weighing Scale...

Страница 20: ...n setting To guarantee the preciseness dead weight calculation will be operated twice 5 3 5 Inputting Max Output Rated Output Voltage mV Input the output value of load cell After mV displayed Check th...

Страница 21: ...key to complete the calibration process In case of multiple pieces of load cells are connected the rated output will be same as single load cell s Because plural load cells are connected with paralle...

Страница 22: ...etting Starting F FUNCTION Mode Press kye 4 times When SETUP is displayed press key 101 00 Input the set value and press key Whenever press key function No is increased Enter the function No that you...

Страница 23: ...kg 01 g 02 ton 111 Language 00 00 Korean 01 English 201 EMPTY Range 100 00 999999 202 Auto Zero Range 00 00 99 Unit 0 25gradation 203 Steady Range 08 01 99 Unit 0 25gradation 204 Steady condition che...

Страница 24: ...Se 01 Disuse 222 Under UNPASS OVERLOAD state Weight display 00 00 Display 01 No display 233 External Input 1 Setting 01 00 Disuse 01 Zero 02 Tare 03 Tare removal 04 Tare Tare removal 05 Hold 06 Hold r...

Страница 25: ...ection under command mode 00 00 Disuse 01 Use 305 Data Format under Stream Mode 00 00 Format1 01 Format2 02 Format3 03 Format4 306 Date transference under stream mode 00 00 Countinuously 01 Single tim...

Страница 26: ...t 03 When input print key 352 Print Format Setting 00 00 Continuous Print 01 Single Print 354 Print Output Delay Time Setting 00 00 10 Unit 1sec 355 Paper Withdraw after printing out Continuos print o...

Страница 27: ...04 Manual Auto At every steady states 05 Manual Auto At the first steady states Display Up Date Speed 104 09 01 09 01 Slow 1 time per 1 sec 09 Fast 60 times per 1 sec Buzzer sound External input detec...

Страница 28: ...eration 207 00 Always active 01 Active under steady condition only Zero key Operation Range 209 00 Active within 2 of Max Capacity 01 Active within 5 of Max Capacity 02 Active within 10 of Max Capacit...

Страница 29: ...Use Unit 1 sec Hold mode 216 00 Sample Hold Hold current weight until Hold Reset 01 Peak Hold Measure Max weight value and hold on display 02 Average Hold Hold average value Hold delay time 217 00 00...

Страница 30: ...05 Hold 01 Zero 06 Hold Hold removal 02 Tare 07 Hold Hold removal 03 Tare removal 08 Print 04 Tare Tare removal 09 Subtotal print External Input 4 Setting 236 00 Disuse 05 Hold 01 Zero 06 Hold Hold re...

Страница 31: ...rmat selection Refer chapter 6 1 4 305 00 Format 1 19byte 01 Format 2 22byte 02 Format 3 17byte 03 Format 4 22byte Stream mode Data transference 306 00 Continuously 01 Single time on every steady stat...

Страница 32: ...Command mode Check Sum detection selection Option Port 311 00 Disuse 01 Use Print Format 352 00 Continuous Print 01 Single Print Print Output Delay Time 354 00 00 09 Unit 1 sec Paper Withdraw Rate se...

Страница 33: ...Serial Number Check H03 S W Version Check Program version check H04 H W Version Check H05 DATE Y M D Check Modification H06 TIME H M S Check Modification 24Hours H07 Password Setting 4digit 1 Tare Ke...

Страница 34: ...0 Manual Current weight Current weight 01 Auto At every steady states Recent Stable weight Stable weight Stable weight 02 Auto At the first steady states Recent Stable weight Stable weight Stable weig...

Страница 35: ...devices Press key 4times When SETUP is displayed Press key to enter Test mode 1 Press key to enter Test mode 2 Press key to go back or cancel Key Test Mode 1 Test Mode 2 Analog Deviation Check Extern...

Страница 36: ...k Mode You can check the deviation of analog value Display current analog value to 0 Display from 1 million unit Display from ten million unit 5 5 2 Display Check Mode 5 5 3 Key Input Check Mode Key o...

Страница 37: ...ue actual output will be 100 of that range EX When analog output mode is 4 20mA 4mA will be out if you enter 4 0 When analog output mode is 4 20mA 20mA will be out if you enter 20 0 When analog output...

Страница 38: ...her devices through serial interface and check the transference and receipt At the normal operation LED will blink one time If you press any key except F testing protocol will be transferred TESTING P...

Страница 39: ...39 SI 460A DIN SIZE WEIGHING INDICATOR 6 INTERFACE 6 1 Serial Interface 6 1 1 RS 422 6 1 2 RS 485 32 pcs of Indicators are connectable 32 pcs of Indicators are connectable...

Страница 40: ...WEIGHING INDICATOR 6 1 3 RS 232 Serial communication interface is sensitive to electric noise Install isolated place from Power cable or other electric cables and wires and please use shielded cable f...

Страница 41: ...r F function 305 00 19byte Header1 Header2 OL OVER LOAD NT NET WEIGHT Tare is not set ST STEADY GS when setting TARE US UNSTEADY 2 Data Format2 ID Number Data Transference Refer F function 101 305 01...

Страница 42: ...e ID Number Refer F function 305 03 22byte LAMP DISPLAY Bit 0 Bit 1 Bit 2 Bit 3 Bit 4 Bit 5 Bit 6 Bit 7 ZERO 1 TARE Gross Weight Print HOLD 1 STEADY 1 Header1 Header2 O OVER G Gross weight S STEADY N...

Страница 43: ...eight STX ID RCWT ETX 22 byte Indicator memory data STX ID RCWD ETX 46 byte Grand Total Data STX ID RGRD ETX 28 byte Current Time Data STX ID RTIM ETX 14 byte Current Date Data STX ID RDAT ETX 14 byte...

Страница 44: ...unit 2byte ETX State1 O Over Load S Steady U Unsteady State2 N Net weight G Gross weight Ex Steady S TARE not used N 0 000kg Indicator memory data ASCII STX ID 2byte RCWD ETX HEX 02 30 31 52 43 57 44...

Страница 45: ...Accumulated Weight 10 000kg Current Time data ASCII STX ID 2byte RTIM ETX HEX 02 30 31 52 54 49 4D 03 Response STX ID RTIM Current Time 6byte ETX 12 00 00 Current date data ASCII STX ID 2byte RDAT ETX...

Страница 46: ...X ID 2byte WHRS ETX HEX 02 30 31 57 48 52 53 03 Response Normal STX ID ACK ETX Error STX ID NAK ETX PRINT Data will be transferred to the port which is set as print mode Function 303 310 02 ASCII STX...

Страница 47: ...Calculate Check sum Sum the value from STX to ETX and converts to ASCII 2byte and transfer Convert the Sum value HEX to ASCII and transmit 28byte ex The sum HEX value from STX to ETX 02 30 31 52 43 5...

Страница 48: ...dbus Function Codes 03 0x03 Read Holding Registers 04 0x04 Read Input Registers 06 0x06 Write Single Registers 16 0x10 Write Multiple Registers CRC Check Method is CRC 16 Add ress Leng th Fea ture Des...

Страница 49: ...Length 2 Lamp data is indicated by 32bit 0 1 2 3 4 5 6 Steady Zero Tare Hold TxD RxD F 3 Error Register Address 21 Length 2 Error data is indicated by 32bit 0 1 Loadcell Error Over Load 4 Standard Key...

Страница 50: ...tput Setting mode Step3 Adjust the displaying value of indicator with keys by 9 999 to make Digital multi meter s value as minimum ex 4mA and save When the SI 460C indicate about 5 200 the Digital mul...

Страница 51: ...How to adjust analog output value Step1 Connect Digital multi meter to the Indicator A out terminal Step2 Enter F function HF14 Minimum Analog Output Setting mode Step3 Adjust the displaying value of...

Страница 52: ...nting format is already programmed and fixed with SE7200 7300 model 6 4 1 Printing Format Using the RS 485 or 422 interface please use convertor and converts to RS 232 and connect with Serial printer...

Страница 53: ...5 3 Isolate Resistance is more than 100M Weight Value is increased regular rate but not return to Zero 1 Load cell Error 2 Load cell connection Error 1 Check Load cell connection 2 Measure Load cell...

Страница 54: ...apacity Eerr06 Amp Gain is too big Check standard weight s weight with set value If there is difference between set value and real weight please re input the value set value is too small Eerr07 Amp Ga...

Страница 55: ...icator s A D with the other indicator 4 Check the power and connection of terminal UNpa55 1 Power is ON when some materials are on weighing part Under Normal Mode if there are more than 20 loading of...

Страница 56: ...Sewhacnm s permission Any kinds of Mal function Defection or External damage caused by operator Any kinds of Mal function Defection caused by using spare part from Non Authorized Distributor or Agent...

Отзывы: