- 15 -
SI 460A DIN SIZE WEIGHING INDICATOR
Indicator will search “DEAE weight” during
10~20 secs automatically to find the best
condition.
※
Over than 1/10,000 resolution setting,
To guarantee the preciseness, DEAD weight calculation (CAL00~CAL09) will be operated
twice.
In this step, if there is unstable condition such as some forces or Vibration on the
scale part, “ErrorA” will be displayed, and “DEAD value” will not be calculated.
Please remove the cause of the force or vibration and process it again.
5-2-6. Calculating span value
If “SPAN” is displayed, input the weight of your “Test Weight” and press
key.