Parallel Print Port selection
●
0
Parallel Port is not installed.
1
Share Standard Serial Port.
F50
2
Share Optional Serial Port.
※
If you install “Print interface option(Centronics Parallel), you have to choose, which port you will
use for print data transference to “Print Interface”.
If you installed “Print interface”, but you didn’t choose the port, you can not print out through
Parallel print.
※
If there is an mal-function on Parallel print device,
“PrinT-Er”
will be displayed.
Then, please check the Parallel print device condition.
Function / Clear key Activation display selection
●
0
Activation display not use
F51
1
Activation
display
use
※
Under F51-01 setting, whenever you press “Function” / “Clear” key, “F” or “C” display will be
appear on main display window, and let you know, “Function” or “Clear” key is activated.
Communication Interval Setting
●
0
Fast Speed (The interval is short)
F52
1
Low Speed (The interval is long)
※
This setting is for “Communication Interval” with P.C or PLC.
If the “Fast Speed” interval is too fast to meet with P.C or PLC, please set “Low Speed” interval.
■
Other Setting
※
Under “Other setting mode”, you can not move to other function directly.
Press
key and move to F01 and move to other function No. directly.
EMPTY Range setting
F80
X.X.X.X.X.X.
(0.0.0.0.1.0)
You can set “EMPTY” Range.
Within set range, indicator will not display current weight and just
display “Zero”.
“0.000” setting : When Net Zero, “Zero” status lamp and Near Zero
relay will be output.
“0.190” setting : Within 190, “Zero” Status lamp and Near Zero relay
will be output.
※
The “EMPTY” range is larger than Max. Capacity, the Error will be displayed.