- 24 -
and move to next step.
Caution
※
(Max. capacity value / division value)
can not be over
20,000
.(as Indicator resolution is 1/200,00).
If the value is over 20,000, Error message
“ Err 01 “
will be displayed and move back
“CAPA”
mode
again.
But, under “Simulation Calibration Mode”, please make (Max. capacity value / division
value)
setting is less than 3,000.
Step No.4 - Measure the “DEAD Weight of Weighing Scale.
Enter the “Dead weight” check mode
Press Enter key.
Indicator will seek optimal Dead Weight automatically.
Under this step, measure the “DEAD Weight of Weighing Scale”.
This “DEAD Weight” is very important to make
“ZERO”
value of weighing scale, so please make sure
that the weighing scale is empty and free from other external variations.
When you press
key, the indicator start measurement and find optimal “Dead weight value
of Scale” automatically.
Caution
※
At this step, if there some force or Vibration on Weighing scale, and unstable condition will be
continued,
“ErrorA”
will be display, and “DEAD value” will not be calculated.
Under this condition, please remove force or vibration and process it again.
Step No.5 - Input Max. Output Rate(mV) of Load cell
Input Load cell
Output Rate(mV)
(refer the load cell label)
Check Max. Rated Output value of
Load cell (Refer the Load cell Label)
After input Load cell’s Max. Rate
output value(mV) and press Enter