44
│ 6 Faults
6
Faults
If a fault occurs during operation, an error message will appear
on the screen.
Error messages are displayed in the order in which they occur.
Up to five errors can be displayed. Error messages continue to
be displayed until the error is corrected.
Overview of possible error messages
Error
code
Error message on
the display
Error correction
9
No calibration
IR sensor adjustment
Adjustment CH4
or
Adjustment
CO2
or
Adjustment gas mixture
measuring
10
Adjustment failed
Test gas
Check test gas
52
XFLASH
SEWERIN Service
Error can only be corrected by
SEWERIN Service
59
Error unknown
SEWERIN Service
Error can only be corrected by
SEWERIN Service
62
IR sensor
Error can only be corrected by
SEWERIN Service
100
Pump error
Probe/filter
Check all filters, probes and hose
connections for porosity and dirt
105
Pump error
Gas outlet
Check pressure at gas input/gas outlet
200
I2C HOST – IR
SEWERIN Service
Error can only be corrected by
SEWERIN Service
201
I2C HOST – EC
SEWERIN Service
Error can only be corrected by
SEWERIN Service
202
I2C HOST – EX
SEWERIN Service
Error can only be corrected by
SEWERIN Service