20
Adjusting the sensitivity
place the test head on the detector's
sensor head
the NOMINAL value of 1.00 VOL.% is
displayed
hold down the release key on the test set
until the indicated concentration (e.g.
TRUE value of 0.85 VOL.%) has
reached a stable value
the nominal value (3 bars) and true value
(1 bar) appear alternately
use the adjusting pin on the service key
(item 6) to clear the attainment of the
maximum indicated concentration
☞
Only now may the test-gas feed be interrupted !
when the adjustment has been
completed (zero point and sensitivity)
the data are stored and the detector
switches automatically to measuring
operation
the concentration-dependent signal
sounds once more