58
│ 6 Maintenance
6.4.6
Required accessories
The following accessories are required for testing the indication
accuracy or the bump test:
●
Test gas
Information about the utilisable test gases can be found in
●
Test set
SPE AutoFlow
6.4.7
Perform a device inspection
6.4.7.1 Requirements (summary)
The following requirements must be met for a device inspection:
−
The device is switched on in clean air.
−
The device continues to draw in clean air after it is switched on.
−
The sensors are warmed up.
−
The necessary accessories are available.
−
The intended test gas is suitable for the gas to be tested.
−
The intended test gas corresponds to the set concentrations
1
.
6.4.7.2 Start device inspection
The device is in measuring mode.
1. Press the Menu key.
2. Select
Device inspection
.
3. Select an application and thus also the variant of the device
inspection (
Warning ECO
,
Warning
,
Measuring
or
Struc-
ture
).
4. Perform the tests described below (section 6.4.7.3 to sec-
1
Setting using the
GasCom
software.