3 Operation
│
25
3.3.3
Switch between levels
3.3.3.1 Changing from measuring mode to menu or measuring
menu
A
WARNING!
Risk to life due to absence of alarms
The device emits alarms only when in measuring mode.
As soon as you switch from measuring mode to the menu,
no alarms are triggered.
●
Switch from measuring mode to the menu only in envi-
ronments that are not explosive and where there is no
toxic or low-oxygen atmosphere.
Switch to the menu
The device is in measuring mode.
●
Press the Menu key. The device switches to the top
menu level.
Switch to the measuring menu
The device is in measuring mode.
●
Press the Enter key. The device switches to the
measurement menu.
3.3.3.2 Switch from menu or the measuring menu into measuring
mode
The display shows the top level menu or the menu (Fig. 7).
●
Wait 5 seconds without pressing a key. The device will auto-
matically switch to measuring mode.
OR
●
Using the arrow keys, select an application and press the En-
ter key.