│ 6 Maintenance
The results can be saved in the device and read out on a com-
puter using the
GasCom
software.
The device reminds the user in good time when the device in-
spection is due.
To perform a guided device inspection, it must be activated us-
ing the
GasCom
software (
Device
>
Edit settings
> Register
General II
).
Note:
In addition, these operating instructions describe only the guided
device inspection.
The visual inspection of probes and probe hoses is not explicitly
part of the guided device inspection. This visual inspection must
therefore also be performed.
6.4.2
Frequency and due date
Device inspection
The frequency of device inspections depends on the application
and the national regulations. The German regulations require
Application Indication accuracy
Acceptance test
(standard device
inspection)
(ECO device
inspection)
Warning
quarterly
before each work shift
Measuring quarterly
quarterly
Structure
weekly
weekly
When switching on, the device draws attention to any device
inspections that are due (Fig. 10). The symbol appears in the
status bar in measuring mode (Fig. 40).