3 Operation
A
CAUTION!
Health risk due to late alarm
If you change the device from the
Warning ExTox
appli-
cation to another application, data that could initiate an
STEL or LTEL alarm is reset to zero.
●
Only switch from
Warning ExTox
to another application
once you have left the work area being monitored.
3.2.11 Settings
You can change the device settings and access information about
the device under
Settings
in the menu (see section 3.3
).
3.2.12 Start/stop/save a measurement
Depending on the application, measurements will have to be
saved or started and then stopped. The ethane analysis is an
exception (see section 3.2.6.3
Application
Measurement
start/stop
save
Inspection above ground
×
Plants
×
Measuring in bar holes
×
Structure
×
Gas measuring
×
Warning %LEL
×
Warning ExTox
×
Difference between starting/stopping and saving
Selecting
Start measurement
followed by
Stop measurement
saves a measurement plot.
Selecting
Save measurement
saves an individual measurement,
the current one.