93
11 Technical aspects
11.3
Error messages
the
EX-TEC Combi
detects
faults itself and displays an
error code in the LCD
Error code Cause, Remedy and Error properties
F1 ................ Sensor error:
No sensor found
Remedy :
switch the device off/on again,
contact SEWERIN-Service
Error properties: the device switches off
F17 .............. Adjustment error:
Zero point of catalytic combustion sensor (CH
4
,C
3
H
8
,
C
9
H
20
),
Remedy:
check test gas, repeat the adjust-
ment, contact SEWERIN-Service
Error properties: 3s alarm, self-resetting
F18 .............. Adjustment error:
Sensitivity of catalytic combustion sensor (CH
4
,C
3
H
8
,
C
9
H
20
),
Remedy:
check test gas, repeat the adjust-
ment, contact SEWERIN-Service
Error properties: 3s alarm, self-resetting
F19 .............. Adjustment error:
Zero point of thermal conductivity sensor (CH
4
,C
3
H
8
),
Remedy:
check test gas, repeat the adjust-
ment, contact SEWERIN-Service
Error properties: 3s alarm, self-resetting
F20 .............. Adjustment error:
Sensitivity of thermal conductivity sensor (CH
4
,C
3
H
8
),
Remedy:
check test gas, repeat the adjust-
ment, contact SEWERIN-Service
Error properties: 3s alarm, self-resetting
Содержание EX-TEC Combi
Страница 1: ...Operating Instructions ...
Страница 3: ...Illustration EX TEC Combi EX TEC Combi Device variation 1 Device variation 2 EX TEC Combi 100 ...
Страница 5: ...Operating Instructions 15 06 2005 V2 X 103401 en EX TEC Combi ...
Страница 117: ...108 Appendix ...
Страница 118: ...109 Appendix ...
Страница 119: ...110 Appendix ...
Страница 120: ...111 Appendix ...
Страница 121: ...112 Appendix ...
Страница 122: ...113 Appendix ...
Страница 123: ...114 Appendix ...
Страница 124: ...115 Appendix ...
Страница 125: ...116 Appendix ...
Страница 126: ...117 Appendix ...