35
5 Testing / maintenance
5.3
Test gases
The following test gases are used in conjunction with the
test set
SPE VOL
tester to test zero point and sensitivity:
Gas
Test gas
Methane CH
4
zero point:
-
fresh air
sensitivity 1:
-
2.2 vol.% (50 %LEL) CH
4
in synth. air
sensitivity 2:
-
100 vol.% CH
4
Propane C
3
H
8
zero point:
-
fresh air
sensitivity
-
1.0 vol.% (59 %LEL) propane (C
3
H
8
)
in synth. air
Nonane C
9
H
20
(
⊗)
zero point:
-
fresh air
sensitivity
-
replacement test gas 0.30 vol.%
propane (C
3
H
8
) in synth. air (reading to
be a returned when replacement test
gas: 0.35 vol.%)
Carbon dioxide CO
2
zero point:
-
fresh air
sensitivity
-
2.0 vol.% (CO
2
) in synth. air
Oxygen O
2
zero point:
-
100 vol.% methane (CH
4
)
sensitivity
-
fresh air
Hydrogen sulphide H
2
S
zero point:
-
fresh air
sensitivity
-
40 ppm hydrogen sulphide (H
2
S) in
synth. air
Carbon monoxide CO
zero point:
-
fresh air
sensitivity
-
40 ppm carbon monoxide (CO) in
synth. air
Ammonia NH
3
(
⊗
)
zero point:
-
fresh air
sensitivity
-
50 ppm Ammonia (NH
3
) in Nitrogen
Содержание EX-TEC Combi
Страница 1: ...Operating Instructions ...
Страница 3: ...Illustration EX TEC Combi EX TEC Combi Device variation 1 Device variation 2 EX TEC Combi 100 ...
Страница 5: ...Operating Instructions 15 06 2005 V2 X 103401 en EX TEC Combi ...
Страница 117: ...108 Appendix ...
Страница 118: ...109 Appendix ...
Страница 119: ...110 Appendix ...
Страница 120: ...111 Appendix ...
Страница 121: ...112 Appendix ...
Страница 122: ...113 Appendix ...
Страница 123: ...114 Appendix ...
Страница 124: ...115 Appendix ...
Страница 125: ...116 Appendix ...
Страница 126: ...117 Appendix ...