23
CREATE CUSTOM TEST SETUP
This test setup page allows custom entry of target pressure, target output, or target %
FS (Full Scale) to be stored for a test by selecting “Custom” from the PROFILE TYPE drop-
down. To change, double click value in any field, and enter a new value. The remaining
fields will update automatically. To accept changes, press “OK”. To ignore changes, press
“Cancel”.
To edit a field, highlight value and type in new value. (Diagram 14)
6.41 TEST SETUP: DIAL GAUGE
Differences from General (Diagram 13)
-Output column removed from test preview window.
6.42 TEST SETUP: P-SWITCH
Differences from General (Diagram 13)
-Output column changes to O/C column to indicate expected switch status (OPEN/
CLOSE).
-Test (Profile ID) selection locked out.
-Profile type selection locked out.
-Test point # entry window locked out.
-Leak Test selection locked out.
Diagram 14
97.00
Custom