49
Expressed in %FS as constant temperature.
Accuracy as RSS non-linearity, hysteresis and non-repeatability
ROOM SUM SQUARES (RSS)
NON-LINEARITY (±0.1%)
2
0.01%
HYSTERESIS (±0.05%)
2
0.0025%
NON-REPEATABILITY (±0.02%)
2
0.0004%
0.0129%
√0.0129%
±0.11%FS at constant temperature
The ability of a transducer to reproduce output readings when the same pressure value
is applied to it consecutively, under the same conditions, and from the same direction.
Non-Repeatability
Hysteresis
The maximum difference in output at any pressure value within the specified range, when
the value is approached with increasing and decreasing pressure.
Long-Term Stability
The ability of a transducer to reproduce output readings obtained during its original cali-
bration at room conditions for a specified period of time.