8.0 CALIBRATING/TESTING A TRANSDUCER/TRANSMITTER
MANUAL CALIBRATION (Typical):
1. Select UUT Profile from either drop-down on real-time screen or by going to the UUT tab
from real-time screen drop down or from UUT tab. Select UUT ID from list and verify.
2. Select REAL TIME tab (If on UUT tab)
3. Connect UUT according to UUT Connection Guide (See Section 7.0).
4. Press the -Setpoint button and select -FS. It will generate the lowest pressure of the UUT
range, wait until pressure is achieved.
5. At this point, you may adjust the zero adjustment on the UUT until the output is within
desired tolerance.
6. Once zero is adjusted, press the Setpoint button and FS button (top of drop-down).
Target pressure will again update in window, wait until pressure is achieved. You may
adjust the span potentiometer until within desired specification.
7. Unit should be calibrated at this point. To verify, select TEST tab.
AUTOMATED TEST:
1. Select RUN TEST tab
2. Select UUT from list, (if not available, follow instructions for creating new).
3. If test profile is not desired, select NEW from list.
4. Press RUN button and wait for completion.
EXPERT SYSTEM TEST:
(See page 37)