LISST-SL V2.1
User’s Guide
30
For cleaning of the windows inside the test section of the
LISST-SL, see Appendix C.
3. Storage
Put plastic caps on the intake and pitot tube openings.
Place the LISST-SL on its foam stands or inside the
shipping box.
Before storage, be sure to turn off power on the TCB.
Extended storage may cause weakening of the battery.
You should recharge the battery before use if the TCB
has not been used for long (month or more).
If preparing to store the instrument system for a long
duration, it is a good idea to run chlorinated water through
the system to prevent biological growth on internal
surfaces. The same is recommended with the filter
provided for Background file collection.
The 5 pin connector on the LISST-SL sensor and the
rectangular connectors that mate to the TCB should be
dressed with a silicone lubricant spray to make insertion
easier and to prevent damage to the rubber.
A 5 pin Impulse plug and sleeve is provided to protect the
connector on the LISST-SL when the sensor is not in use.
The 3 pin Impulse plug should be put on the TCB
connector
and
the
rubber
cover
put
on
the
communications connector to keep the connections dry
and clean.