TEMPERATURE SLOPE CORRECTION 

When creating or editing a configuration file, temperature slope correction can be 
applied. Temperature slope correction is used to remove any length dependency of 
the temperature values for optical fibre. Length dependency manifests itself as a 
gradual slope in the distributed temperature sensing trace over the length of a 
uniform area along the optical fibre. Temperature slope correction is normally 
performed by entering a predetermined fixed value or by referencing two regions of 
the optical fibre that have a uniform temperature. The exact temperature of the 
region is not required. The Sensornet DTS software automatically calibrates the 
trace as required. 

TEMPERATURE OFFSET 

Temperature offset allows the user to calibrate the absolute temperature values for 
the Sentinel DTS. This is normally performed by entering a predetermined fixed 
value or by referencing a section of the optical fibre in a specific region with the 
temperature values of a reference probe or region of known temperature. 

CHART RECORDER POINTS 

Chart recorder points are used to monitor specific points of interest along the optical 
fibre. Once chart recorder points are created they can be viewed in real time while 
measurements are being taken. The values are displayed as temperature vs time. 

TRACE AVERAGING 

Trace averaging allows the user to specify the measurement time and amount of 
spatial averaging applied. The measurement time is the amount of time the Sentinel 
DTS averages distributed temperature sensing data before displaying it on the 
graphical display. By increasing the measurement time, better temperature 
resolution can be achieved. Spatial averaging is used to remove any unwanted 
random variability or noise from the trace. Increasing the value of spatial averaging 
can remove noise and smooth out the trace but this may also remove useful 
information. Unless the desired temperature resolution cannot be obtained by 
increasing the measurement time, Sensornet generally recommends leaving the 
spatial averaging on the default measurement setting. Both measurement time and 
spatial averaging are applied by entering a predetermined value or adjust the trace 
using the sliders. 

 

Содержание Sentinel DTS

Страница 1: ......

Страница 2: ...This page intentionally blank ...

Страница 3: ...nching the Sensornet Sentinel DTS Software 7 5 THE SENSORNET DTS SENTINEL DATA DISPLAY SCREEN 9 The Temperature Sensing Trace Display Field 9 The Trace Selection Field 10 The Spatial Averaging Adjustment Control 10 The Data Recording Information Field 10 The Program Control Field 10 6 CREATING CONFIGURATION FILES 11 Length Correction 11 Measurement Range 11 Temperature Slope Correction 12 Temperat...

Страница 4: ...Page ii Sentinal DTS Quick Start Guide SEN2 QSG1 0 This page intentionally blank ...

Страница 5: ...re that these warnings are read and understood at all times NOTES Used throughout the user guide to provide additional information hints and tips WARNING THE SENTINEL DTS IS HEAVY ENSURE ADEQUATE CARE IS TAKEN WHEN LIFTING AND MOVING THE SENTINEL DTS THE SENTINEL DTS SHOULD BE MOUNTED ON A SUITABLE SURFACE OR RACK MOUNT CAPABLE OF SUPPORTING THE SENTINEL DTS WEIGHT THE SENTINEL DTS WEIGHS 23KG WAR...

Страница 6: ...N Never operate the Sentinel DTS if any damage is observed Contact Sensornet Ltd for advice CAUTION Do not expose the Sentinel DTS to adverse temperatures This may seriously interfere with the correct operation of the Sentinel DTS LASER SAFETY The Sentinel DTS is a Class 1M laser product The Sentinel DTS distributed sensing system incorporates a pulsed laser device whose invisible output radiation...

Страница 7: ... the following safety warnings cautions and notices should be observed WARNING DO NOT LOOK AT OR EXAMINE EXPOSED FIBRE OR CONNECTOR ENDS WHENEVER THE DTS IS IN OPERATION WARNING AVOID EYE EXPOSURE TO THE INVISIBLE LASER RADIATION EMITTED FROM FIBRE CONNECTORS OR EXPOSED FIBRE ENDS WARNING DO NOT EXAMINE OR INSPECT EXPOSED FIBRE OR CONNECTOR ENDS USING A VIEWING AID SUCH AS AN EYE LOUPE OR MICROSCO...

Страница 8: ...de SEN2 QSG1 0 2 CONNECTING THE SENTINEL DTS MULTIPLEXER The following connections should be made between the Sentinel DTS and if required the Multiplexer and the installation Figure 1 The Sentinel DTS and Multiplexer Connection Diagram ...

Страница 9: ...ide SEN2 QSG1 0 Page 5 For more information on making connections using optical fibres please refer to the Sentinel DTS User Guide 3 THE SENTINEL DTS CONTROLS AND INDICATORS THE SENTINEL DTS Figure 2 The Front of the Sentinel DTS ...

Страница 10: ...d is emitting LASER ENABLE SWITCH The Laser Enable Switch is a safety device and is used to manually prevent laser operation To enable laser operation turn to the horizontal position To disable laser operation turn the switch to the vertical position CD RW DRIVE The CD RW drive can be used to load and save configuration and measurement data FLASH DISK READER WRITER The Flash Disk Reader Writer can...

Страница 11: ...TS system by switching the power isolation switch on the rear panel to the ON position The Sentinel DTS is configured to automatically load and run the DTS measurement software However if the software is not running once the operating system has loaded double click on the Sentinel DTS Icon as shown in Figure 3 or alternatively click on Start Programs Sentinel DTS Figure 3 The Sensornet DTS Softwar...

Страница 12: ... 4 The Sensornet DTS Sentinel Main Menu To begin measurements immediately using the factory set calibration values click Start Measurement Alternatively the configuration file can be modified using either the Configuration Wizard or the Configuration Editor ...

Страница 13: ...lay screen is shown in figure 5 Figure 5 The Sensornet DTS Sentinel Data Display Screen THE TEMPERATURE SENSING TRACE DISPLAY FIELD The Temperature Sensing Trace Display Field shows the distributed temperature trace of the optical fibre being measured in a graphical format The default display type is Distributed Temperature which displays the trace data as temperature vs distance in the selected u...

Страница 14: ...ial averaging applied to the distributed temperature trace data Apply spatial averaging to remove unwanted noise but at the expense of degrading spatial resolution THE DATA RECORDING INFORMATION FIELD The Data Recording Information Field displays information regarding the distributed temperature sensing data determined by the configuration file including start and stop times the current repetition...

Страница 15: ...r Run Configuration Wizard The following parameters can be set when creating configuration files LENGTH CORRECTION When creating or editing a configuration file length correction can be applied This is used to correct for dispersion within the fibre cable and ensure both the stokes and anti stokes data are range calibrated This is normally performed by referencing to the end of the optical fibre a...

Страница 16: ...alues of a reference probe or region of known temperature CHART RECORDER POINTS Chart recorder points are used to monitor specific points of interest along the optical fibre Once chart recorder points are created they can be viewed in real time while measurements are being taken The values are displayed as temperature vs time TRACE AVERAGING Trace averaging allows the user to specify the measureme...

Страница 17: ...ution The repetition time is the total time between consecutive temperature sensing measurements and may include some idle time if the repetition time is greater than the measurement time The Sentinel DTS may be set to operate continuously by checking the appropriate tick box or for a limited number of repetitions un checking the tick box and entering the desired number of repetitions For more inf...

Страница 18: ... Expanded Directory Route The Sentinel DTS generates two types of distributed data files These are Full Data and Temperature Only data files FULL DATA SET FILES Full Data Set files contain the complete data set generated by the Sentinel DTS system after every repetition of the measurement cycle These contain the distributed temperature data raw signal data and various installation details name tim...

Страница 19: ...5 TEMPERATURE ONLY DATA SET FILES The Temperature Only Data Set files contain only the temperature data generated by the Sentinel DTS system after every repetition of the measurement cycle The Temperature Only Data Set files have a dtd file extension ...

Страница 20: ... Temperature trace exhibits step changes at fibre joints Dirty connectors Large splice losses Different fibre types joined Clean connectors Re splice fibre joints Unstable calibration temperature trace jumps excessively after each measurement Incorrect slope or offset on temperature trace Calibration options defined inappropriately Fibre loss variations e g bad connector or fibre join Check refere...

Страница 21: ...k router or gateway settings with network administrator Reset VNC passwords Multiplexer channel data appears incorrect Multiplexer Expansion Unit error Fibre cables connected to wrong multiplexer channels Expansion unit cable fault Check fibre cable connections and active channels Ensure Expansion unit cable is connected correctly If problem persists contact Sensornet Ltd DAQ timeout error Interlo...

Страница 22: ...Page 18 Sentinal DTS Quick Start Guide SEN2 QSG1 0 This page intentionally blank ...

Страница 23: ...This page intentionally blank ...

Страница 24: ...Sensornet Ltd 198 Providence Square Jacob Street London SE1 2DZ United Kingdom Tel 44 0 20 7394 3555 Fax 44 0 20 7394 0555 E mail enquires sensornet co uk Website www sensornet co uk ...

Отзывы: