Transducer Control 5 User Manual (RWT3746IM)
Revision 7, September 2020 - Page 19
Digital Setup
The
“Digital Setup” tab of the User Configuration allows the user to control some of the
operational parameters that effect the processing and output of data.
Auto Reset
The Auto Reset settings control the Peak Torque Auto Reset value. Auto reset is a
method of storing a peak value until a reset condition is met. This is especially useful
in machines that cycle, e.g. bottle top tightener.
The reset parameters control the reset point, hold time and when to apply the reset.
Depending on your transducer firmware version, some of these settings may not be
available. The Auto Reset capabilities were upgraded in version 5.2.0 of the firmware.
With each of the settings, move the slider to the value required. The settings are
described below.
Reset Percentage
The reset percentage controls the point at which the Peak Torque Auto Reset value
will hit its reset trigger. The reset point is a percentage of the stored Peak Torque Auto
Reset value, compared against the current torque value.
Reset Hold Time
The reset hold time is the amount of time in milliseconds (ms), which the Peak Torque
Auto Reset value will hold at its peak value for, before it is reset.
Reset Trigger Time
The Reset Trigger Time is the length of time the torque value needs to be below the
reset trigger level, for a reset to be triggered.