Transducer Control 5 User Manual (RWT3746IM)
Revision 7, September 2020 - Page 18
User Defined Voltage
If a custom analog output scaling is required select the
“User Defined Voltage”
option
from the “Voltage Scale Select” combo box, then use the “Zero Offset” and
“Voltage Swing” slider bars to configure the required scaling. The text boxes to the
right of the respective slider bars will display the configured voltage.
Zero Offset
The
“Zero Offset” allows the user to change the zero point voltage between 0 and
9.5V. If the
“Zero Offset” is increased, the maximum 10V “Voltage Swing” voltage
will be reduced.
This feature is useful if the user has an analog capture device that does not support
negative voltages (unipolar only), but requires the direction of torque.
Voltage Swing
The
“Voltage Swing” sets the voltage span of the analog output. The voltage can
be stepped in 0.1V increments and can be scaled between 0.5V and 10V. The
maximum voltage will be reduced if the
“Zero Offset” is not 0V.
Unipolar / Absolute Value
The “Unipolar / Absolute Value” checkbox option switches the analog output into a
positive only voltage. All input values will be made absolute, making all negative
values positive, i.e. -5Nm value will 5Nm. This has the effect of removing
the torque direction from the voltage output.