Operator Manual
Page
32
of
40
P/N 1580281
SEC3120 Digital Transmitter
Sensor Electronics Corporation
Rev 10, 20150707
By selecting ENTER on YES/NO the user may
change the value using the UP/DN arrows. After
selecting ENTER after setting the value, the
operator may move the cursor to PREVIOUS
MENU and select ENTER to go back to the main
menu after choosing to SAVE or ABORT the
changes.
14.
Data Logging (Optional)
The SEC3120 Unit provides event data logging to a flash card. This flash card can be read by any personal
computer that can read a FAT16 format file system, similar to cards from digital cameras and other portable
devices. It should NEVER be formatted by a personal computer, rather be formatted by the SEC3120 unit
itself under the Initial/Top Menu item “Format Flash Card”. Data can be read directly from the flash card or
copied to a personal computer. The file is a text file containing comma separated data values, one event
per line. The text file (“SEC3100.TXT”) will be the only file on the flash card, and currently will NOT be
allowed to grow beyond 16MB (th
is is considered the memory card “full now” state). After the data is
archived from the flash card, it can be re-formatted to start storage over. A flash card will typically not
become full for many years.
Note: You cannot use a memory card formatted by an SEC3120 in an SEC3100 Unit!
You may, under
certain circumstances, use a memory card formatted by an SEC3100, however the SEC3120 will consider
it full when it approaches the SEC3100 file size limitation of 4MB. It is best to reformat such a card to make
more use of its capacity.
As The Memory Card Fills Up
When a flash
card crosses 80% full (approximately 12 MB), the memory card status will change from “MC-
OK” to “MC-FULL”, warning the user to copy the contents off to a permanent storage location and reformat
the card to start storage over. The SEC3120 will continue to log data to the flash card until it reaches 100%
full. At that time, the status will change to “MC-FULLNOW” indicating that there is no more room to store
data. At this point data storage has STOPPED and events may be lost that would otherwise be stored. If
the card is removed and reinserted, the SEC3120 may eventually display the status as “MC-FAULT”,
indicating a memory card fault. The card MUST be formatted at this point.
How Long It May Take To Fill a Memory Card
The absolutely shortest period of time to fill a memory flash data card is approximately 200
– 300 hours.
To accomplish this unreasonable feat gas levels must be constantly changing beyond 5% of sensor range
and alarm events and other related events must be persistently changing at a highly sustained rate. It is
unlikely that this could ever happen under normal circumstances, since alarms would be sounding and
intervention would be absolutely necessary. If sensor(s) are properly calibrated, and normal maintenance
is performed, the memory card will probably not fill up for ten years or more. Since the operating
environment determines the amount of data and frequency for storage, the time it takes to fill a data card
will differ for each installation.
Flash Card Removal and Formatting
Removing a data flash card should not occur without selecting the first Top Menu Item “Eject Flash Card”.
This prepares the flash card for removal by writing any data cached in memory out to the file system and
ensures the file system on the data flash card is not corrupted. Formatting a data flash card allows the card
to start over and re-capture space. Caution should be exercised to ensure that any data needed is archived
first since this process will erase all data. The file system will be re-started and prepared as if from the
factory. Just select the Menu Item on the second page of the Top Menu “Format Flash Card” to begin the
process.
Data Log File Contents
HIDE LOW GAS:
*
YES / NO *
PREVIOUS MENU
*
SAVE CHANGES
ABORT CHANGES
Содержание SEC 3000
Страница 15: ...Doc 1460003 Rev A Operation Manual SEC 3000 Electro Chemical Gas Sensor Page 12 of 13 ...
Страница 54: ...Part Number 3100 EXP MANUAL REV 11 20141007 FIGURE 4 SEC 3100 SEC Millenium Mounting ...
Страница 55: ...Part Number 3100 EXP MANUAL REV 11 20141007 FIGURE 5 SEC 3100 SEC 3000 Mounting ...
Страница 56: ...TB 1 J1 TB 2 J2 1 6 1 6 1 4 1 6 TB 5 J5 TB 6 J6 ...
Страница 100: ...Operator Manual Page 39 of 40 P N 1580281 SEC3120 Digital Transmitter Sensor Electronics Corporation Rev 10 20150707 ...
Страница 101: ...Operator Manual Page 40 of 40 P N 1580281 SEC3120 Digital Transmitter Sensor Electronics Corporation Rev 10 20150707 ...
Страница 148: ...SEC 3120 Wiring TB 1 P3 TB 2 P6 1 6 1 6 1 2 3 4 1 2 3 4 7 6 5 4 3 2 1 TB 3 P1 TB 4 P5 ...
Страница 156: ......
Страница 158: ......
Страница 159: ......
Страница 160: ......
Страница 161: ......
Страница 399: ...SEC P N 1460014 Rev ECO A 000209 Page 11 of 26 ...
Страница 400: ...SEC P N 1460014 Rev ECO A 000209 Page 12 of 26 ...
Страница 431: ...Part Number 1460005 Revision C ...
Страница 433: ...Part Number 1460005 Revision C ...
Страница 434: ...Part Number 1460005 Revision C ...
Страница 435: ...Part Number 1460005 Revision C ...
Страница 448: ...10 VII Drawing Section Figure Title Figure 1 Wiring Diagram SEC MILLENIUM Figure 2 SEC Sensor Separation Kit ...
Страница 473: ...10 VII Drawing Section Figure Title Figure 1 Wiring Diagram SEC MILLENIUM Figure 2 SEC Sensor Separation Kit ...