Doc#: 1460003, Rev A
Operation Manual, SEC 3000 Electro-Chemical Gas Sensor
Page
7
of
13
Shielded cable is recommended. Wiring should be installed in medal conduit with no other cabling in the
same conduit.
Warm-up
When power is applied to the detector, it enters a one (1) minute warm-up mode. The output current will be
0.8 mA during the warm up time period. At the end of the warm-up period with no faults present, the
detector automatically enters the normal operating mode (4 mA). If a fault is present after warm-up, the
detector current output will indicate a fault. See the following chart for fault code status. Some
electrochemical sensors will take up to 24 hours to stabilize. Newly installed sensors should be calibrated
after they have been allowed to stabilize.
Normal
In the normal operating mode, the 4 to 20 mA signal levels correspond to the detected gas concentration.
The detector continuously checks for system faults or initiation of calibration and automatically changes to
the appropriate mode.
The 4 to 20 mA output of the SEC 3000 is a non-isolated current source.
Current Output and Corresponding Status
Once the fault is cleared the SEC 3000 will automatically resume normal operation.
Current Output
Status
.
0-20 mA
Normal measuring mode
0.0 mA
Unit Fault
0.8 mA
Unit warm up
1.2 mA
Zero drift fault
1.6 mA
Calibration fault
2.0 mA
Unit spanning
2.2 mA
Unit zeroing
4-20 mA
Normal measuring mode
4.0 mA
Zero gas level
5.6 mA 10% Full Scale
8.0 mA 25% Full Scale
12 mA 50% Full Scale
16 mA 75% Full Scale
20 mA
Full scale
>20 mA Over-range
Содержание SEC 3000
Страница 15: ...Doc 1460003 Rev A Operation Manual SEC 3000 Electro Chemical Gas Sensor Page 12 of 13 ...
Страница 54: ...Part Number 3100 EXP MANUAL REV 11 20141007 FIGURE 4 SEC 3100 SEC Millenium Mounting ...
Страница 55: ...Part Number 3100 EXP MANUAL REV 11 20141007 FIGURE 5 SEC 3100 SEC 3000 Mounting ...
Страница 56: ...TB 1 J1 TB 2 J2 1 6 1 6 1 4 1 6 TB 5 J5 TB 6 J6 ...
Страница 100: ...Operator Manual Page 39 of 40 P N 1580281 SEC3120 Digital Transmitter Sensor Electronics Corporation Rev 10 20150707 ...
Страница 101: ...Operator Manual Page 40 of 40 P N 1580281 SEC3120 Digital Transmitter Sensor Electronics Corporation Rev 10 20150707 ...
Страница 148: ...SEC 3120 Wiring TB 1 P3 TB 2 P6 1 6 1 6 1 2 3 4 1 2 3 4 7 6 5 4 3 2 1 TB 3 P1 TB 4 P5 ...
Страница 156: ......
Страница 158: ......
Страница 159: ......
Страница 160: ......
Страница 161: ......
Страница 399: ...SEC P N 1460014 Rev ECO A 000209 Page 11 of 26 ...
Страница 400: ...SEC P N 1460014 Rev ECO A 000209 Page 12 of 26 ...
Страница 431: ...Part Number 1460005 Revision C ...
Страница 433: ...Part Number 1460005 Revision C ...
Страница 434: ...Part Number 1460005 Revision C ...
Страница 435: ...Part Number 1460005 Revision C ...
Страница 448: ...10 VII Drawing Section Figure Title Figure 1 Wiring Diagram SEC MILLENIUM Figure 2 SEC Sensor Separation Kit ...
Страница 473: ...10 VII Drawing Section Figure Title Figure 1 Wiring Diagram SEC MILLENIUM Figure 2 SEC Sensor Separation Kit ...