Part Number 3100-EXP-MANUAL
REV 11 20141007
The 4 to 20 mA output of the SEC 3100 sensor is a non-isolated current source.
Current Output and Corresponding Status
Once the fault is cleared the SEC 3100 will atomatically resume normal operation.
Flash
Rate
Output Current
Unit Status
Label
Possible Problem
1
4-20ma
Unit Running
Unit is measuring gas and adjusting 4-20ma output accordingly.
2
2.2ma
Unit Zero Calibrating
Unit going through its
zero
calibration
procedure.
3
2.0ma
Unit Spanning
Unit going through its
spanning
procedure.
5
0.8ma
Unit Warm-up
Only for one minute after unit power-up
6
0.0ma
Power-up Fault
Hard Fault (refer to gas sensor manual)
7
1.6ma
Calibration Fault
1.
Attempt
Unit Span
with no gas
2.
Attempt
Unit Zero
with gas
8
NA
NA
Currently Not Used
9
0.0ma
Unit Fault
Hard Fault (refer to gas sensor manual)
10
1.0ma
Optics Fault
Clean sensor’s windows
11
1.2ma
Zero Drift Fault
Hard Fault (refer to gas sensor manual)
12
0.0ma
Configuration Fault
Hard Fault (refer to gas sensor manual)
16
0.2ma
Reference Channel Fault
Hard Fault (refer to gas sensor manual)
17
0.4ma
Active Channel Fault
Hard Fault or
Unit Zero
with gas
Current Output
Status
0-20 mA
Normal measuring mode
0.0 mA
Unit Fault
0.2 mA
Reference channel fault
0.4 mA
Analytical channel fault
0.8 mA
Unit warm up
1.0 mA
Optics fault
1.2 mA
Zero drift fault
1.6 mA
Calibration fault
2.0 mA
Unit spanning
2.2 mA
Unit zeroing
4.0 mA
Zero gas level (0% of full scale)
5.6 mA
(10% of full scale)
8.0 mA
(25% of full scale)
12 mA
(50% of full scale)
16 mA
(75% of full scale)
20 mA
Full scale (100% of full scale)
20.1- 23 mA Over-range (> 100% of full scale)
Содержание SEC 3000
Страница 15: ...Doc 1460003 Rev A Operation Manual SEC 3000 Electro Chemical Gas Sensor Page 12 of 13 ...
Страница 54: ...Part Number 3100 EXP MANUAL REV 11 20141007 FIGURE 4 SEC 3100 SEC Millenium Mounting ...
Страница 55: ...Part Number 3100 EXP MANUAL REV 11 20141007 FIGURE 5 SEC 3100 SEC 3000 Mounting ...
Страница 56: ...TB 1 J1 TB 2 J2 1 6 1 6 1 4 1 6 TB 5 J5 TB 6 J6 ...
Страница 100: ...Operator Manual Page 39 of 40 P N 1580281 SEC3120 Digital Transmitter Sensor Electronics Corporation Rev 10 20150707 ...
Страница 101: ...Operator Manual Page 40 of 40 P N 1580281 SEC3120 Digital Transmitter Sensor Electronics Corporation Rev 10 20150707 ...
Страница 148: ...SEC 3120 Wiring TB 1 P3 TB 2 P6 1 6 1 6 1 2 3 4 1 2 3 4 7 6 5 4 3 2 1 TB 3 P1 TB 4 P5 ...
Страница 156: ......
Страница 158: ......
Страница 159: ......
Страница 160: ......
Страница 161: ......
Страница 399: ...SEC P N 1460014 Rev ECO A 000209 Page 11 of 26 ...
Страница 400: ...SEC P N 1460014 Rev ECO A 000209 Page 12 of 26 ...
Страница 431: ...Part Number 1460005 Revision C ...
Страница 433: ...Part Number 1460005 Revision C ...
Страница 434: ...Part Number 1460005 Revision C ...
Страница 435: ...Part Number 1460005 Revision C ...
Страница 448: ...10 VII Drawing Section Figure Title Figure 1 Wiring Diagram SEC MILLENIUM Figure 2 SEC Sensor Separation Kit ...
Страница 473: ...10 VII Drawing Section Figure Title Figure 1 Wiring Diagram SEC MILLENIUM Figure 2 SEC Sensor Separation Kit ...