- 1 0 -
SW No. 5
The transmission output during calibrationscanbespecified 2 types with the ON-OFF
of SwitchNo.5.
ON
1
2
3
4
5
6
Continuous output
Transmission output corresponds to the measuredvalues a s s a m e a s t h e
■
Normal Operations Mode. Therfore, transmission output corresponds to
standard solutionvalues.
Hold just before
Transmission outputisfixedtothemeasuredvalueswhichareatjust
■
before entering into Calibration Mode. Therefore, the transmission output is
kept beingfixedduringcalibrations by standardsolutions.
Afterreturned to theNormal Operation Mode, transmission output is released
and corresponds to measured value.
SW No. 6
This s w i t c h i s o u t o f u s e
3 .Wiring
3 . 1
Wiring of Alarm Contact Output
(
)
Caution
Contact capacity is specified with AC 250V , 3A
resistance load, COS
φ=1.0
When switching the load more than this, use power relay to fit the load connected.
(
)
Caution
Before wiring theload,checkon which control action upper limit/lower limit
ALM1, ARM2 are set.
Pay attention in wiring load t o s o t h a t C O M - B c o n t a c t i s O N when instrument
Caution
power is OFF.
(
)
This CE-8E has 2 channels of alarmcontactoutput;ALM1 and ALM2 1c contact output
Control actionofbothALM1andALM2canbesetupperlimitactionorlower limitaction by
DIP. Switches.
At deliverytoyou,CE-8Eisinitiallysetasfollows;
( )
At delivery to you;
ALM1
Upper Limit Action H
( )
ALM2
Lower Limit Action L
The display shows initial set contents when the instrumentispowered ON.